Bioball universal modular neck adapter as a salvage for failed revision total hip arthroplasty


1 Department of Orthopaedics and Joint Replacement Surgery, Indraprastha Apollo Hospital, New Delhi, India
2 Department of Orthopaedics and Joint Replacement Surgery, Asutosh Hospital, Surat, Gujarat, India

Click here for correspondence address and email

Date of Web Publication17-Sep-2013
 

   Abstract 

Management of recurrent dislocation of total hip arthroplasty is often a challenging and daunting task. Re-revision of such a total hip prosthesis may not be easy as the removal of a well-fixed, fully coated stem is extremely difficult. We managed to salvage instability in three revision hip cases in which the fully coated stem had subsided by using a bioball universal neck adapter without changing the femoral stem or acetabular cup.

Keywords: Failed arthroplasty, bioball, revision total hip

How to cite this article:
Vaishya R, Sharma M, Chaudhary RR. Bioball universal modular neck adapter as a salvage for failed revision total hip arthroplasty. Indian J Orthop 2013;47:519-22

How to cite this URL:
Vaishya R, Sharma M, Chaudhary RR. Bioball universal modular neck adapter as a salvage for failed revision total hip arthroplasty. Indian J Orthop [serial online] 2013 [cited 2019 Dec 11];47:519-22. Available from: http://www.ijoonline.com/text.asp?2013/47/5/519/118211

   Introduction Top


The use of a fully coated stem is an option for revision total hip arthroplasty (THA), considering the scratch fit obtained by it in the diaphysis and later incorporation by bony in growth. [1],[2] However, in the early postoperative period, this prosthesis may sometimes subside before adequate bone incorporation has taken place. [3],[4],[5] This subsidence of the femoral stem may be followed by later bone incorporation with the prosthesis in a sunken and/or rotated position. Short offset or incorrect version of the stem may lead to recurrent dislocations and failure of the revision THA. Re-revision of such a total hip prosthesis may not be easy as the removal of a well-fixed stem is extremely difficult and any vigorous attempt at its removal can lead to femoral shaft fracture. We managed to salvage three hips with subsidence of extensively porous coated cylindrical femoral stem (Solution System, Depuy International Ltd., UK) by using a modular neck extension called the bioball universal neck adapter (Merete, Germany). This alleviated the need to revise any component and the limb length and tissue tensioning were restored to normal.


   Case Reports Top


Case 1

A 70 year old female was operated for left femoral neck fracture with THA using metal on metal prosthesis (ASR). 1 year later she developed aseptic loosening and a peri-prosthetic fracture of the femur, which was revised with fully coated stem (Solution System, Depuy International Ltd., UK) and cerclage wires [Figure 1]a. Ten weeks after the surgery, on weight bearing, the stem subsided and the hip dislocated [Figure 1]b. Because maximum size of the femoral neck was already used in the first surgery, the acetabular liner was changed to a constrained liner. But, after 1 month of the surgery, the neck got dislodged from the stem [Figure 1]c. Version was found changed intraoperatively. This difficult hip with recurrent dislocation was salvaged using the bioball neck adapter and the version was restored. At 2 years followup, the fracture has united and the patient is pain free and mobile, with no further episode of dislocation [Figure 1]d. There was no thigh or trochanteric pain till the last followup.
Figure 1: X-ray anteroposterior view of the left hip joint showing (a) revision total hip arthroplasty using a fully coated stem (b) stem subsidence and dislocation of hip (c) dislodgement of the neck from the stem (d) salvage using a bioball neck adapter.

Click here to view


Case 2

A 62 year old male had cemented THA for idiopathic avascular necrosis, but started to have progressively increasing pain and limping 3 months after the surgery. The X-rays confirmed early and severe osteolysis. This hip was revised by using a fully coated stem [Figure 2]a. The stem showed subsidence in 4 months time [Figure 2]b, despite keeping the patient nonweight bearing for 2 months on crutches. It was managed successfully using the bioball adapter [Figure 2]c, with no incidence of re-dislocation at 1.5 years followup. There was no trochanteric pain or thigh pain.
Figure 2: X-ray anteroposterior view of the left hip joint showing (a) Revision total hip arthroplasty using a fully coated femoral stem (b) Subsidence of the femoral stem (c) Re-revision with a bioball neck adapter

Click here to view


Case 3

A 35 year old male (ankylosing spondylitis) with bilateral hip involvement was operated elsewhere and had bilateral total hip replacement, but both the hips got infected. The infected implants were removed and the patient was left with a bilateral excision arthroplasty (Girdlestone's procedure). After a period of 1 year, after the infection had subsided, he was operated for left hybrid THA. While waiting for the right revision hip arthroplasty, he fractured his right femur after a trivial fall. He was operated on the right side and a revision total hip replacement was performed using the fully porous coated stem (Solution Depuy). After 3 months, when the patient was made to bear weight, the stem subsided and the patient developed a painful right hip. The pain was due to subsidence and instability as the patient could not bear weight on the affected side. The version was found to be changed intraoperatively. We managed to restore the hip biomechanics and restore the version using the bioball neck adapter. The patient has been under followup for 3 years now without any further incidence of instability.


   Discussion Top


Extensively coated stems in revision arthroplasty have shown excellent mid-term results. Chung et al. in their study on Paprosky femoral type III defects in 96 femoral revisions achieved stable bony in-growth in 92 cases. [1] Jaya Kumar followed 56 revision arthroplasty cases using extensively porous coated femoral stem for 6 years. All implants demonstrated evidence of bony in-growth and stable fixation, with no cases of loosening, instability, deep infection, stress shielding, subsidence or osteolysis at last followup. [2] Moon et al. also showed similar results in 35 patients followed up for 77.5 months. [6] Kim et al. used fully porous coated stems with strut allografts in 54 hips followed for 10 years. Only two hips showed aseptic loosening and had to be revised. [7] Similar results have been shown by Whiteside and Engh et al. [8],[9] These authors have thus recommended the use of extensive coated stems in revision arthroplasty.

However, failure of adequate bone in-growth, fibrous in-growth or unstable implant may lead to subsidence. Weeden et al. followed 170 patients for a mean of 14.2 years and reported bone in-growth in 82% of the hips, stable fibrous fixation in 14% of the hips and 4% unstable hips. They concluded that failure of fixation correlated highly with extent of bone loss at the time of surgery. [3] Porous coated stems however have shown good mid-term results, but may have certain limitations. Zhao et al. reported prosthesis subsidence of 5 mm in one case and 7 mm in another case of 20 cases operated with fully porous coated stem and followed for 36 months. [4] The mean time to subsidence was 3.1 months in our cases. This raises a question on the optimal time to weight bearing after fixation with these fully coated cementless stems. We made our patients to weight bear not before 2 months, but still all three stems showed subsidence. The exact time to bear weight is not certain as there are no clinical or radiological parameters to guide the surgeon regarding bone in-growth in this initial period. Bone in-growth in the stem is also often not visible on plain X-rays. Fretting caused by micro-motion between the modular components may lead to particle and ion release to the surrounding tissues. Trunionosis may lead to peri-prosthetic bone loss and third body wear. [5] Only long-term studies can show the effect of the titanium ions released into the surrounding tissues. These coated stems cannot be used in femora with Paprosky type IIIB defects with diameter >19 mm or in type IV defects, [10] and intraoperative fractures are common with large diameter femoral stems. [11]

Although we found the modular neck stem system of bioball to be a great salvage procedure for difficult recurrent dislocation of THA, we believe that the long term outcomes of modular neck-stem connection still needs to be proven in more clinical studies as no previously published reports are available in the literature. The Bioball Adapter System (designed by Merete, a German Company and is distributed by WG Healthcare, UK) seems to be a safe, effective and least traumatic way to salvage instability in revision THA. It allows various alterations to neck length and offset [Figure 3] to enable a stable, simple and quick revision without the need for complicated and costly removal of the existing components. Thus, it allows a surgeon to manage unexpected surgical situations with a high degree of flexibility and precision. This innovative hip endo-prosthetic modular head system appears to be the only such available system at present that helps the surgeon to intraoperatively correct femoral stem malpositioning to minimize the risk of dislocation. It is indicated for neck length equalization and antetorsion correction of the femoral hip stem during revision hip surgery. The main advantages of the bioball system are allowing precise adjustment of neck length, anteversion and offset of the well-fixed femoral component, with a simple instrumentation. The bioball adapters are made from TiAl6V4 according to ISO 5832-3. Regarding the risk of fracture, the bioball adapters are tested with merete hip stems according to ISO 7206-6, the testing standard for testing hip stem necks. All testings were successful. Bioball heads are available in 28 mm, 32 mm and 36 mm metal (vivium) and ceramic (zirconium oxide and Delta Ceramic) and it can give up to 21 mm extra neck length and has 7.5 degrees dialable offset. These can be used on an existing stem to switch to or from a ceramic head with the simple use of a single adapter, thereby avoiding full femoral revision and any unwanted cross-material contamination. A bioball adapter can fit a variety of tapers, including 12/14 and 14/16. The bioball universal neck adapter is perhaps one of the least traumatic ways to deal with the situation as there is no need to change any other component of the total hip system.
Figure 3: Bioball neck adapter system showing variable neck lengths and offsets

Click here to view


 
   References Top

1.Chung LH, Wu PK, Chen CF, Chen WM, Chen TH, Liu CL. Extensively porous-coated stems for femoral revision: Reliable choice for stem revision in Paprosky femoral type III defects. Orthopedics 2012;3:1017-21.  Back to cited text no. 1
    
2.Jayakumar P, Malik AK, Islam SU, Haddad FS. Revision hip arthroplasty using an extensively porous coated stem: Medium term results. Hip Int 2011;21:129-35.  Back to cited text no. 2
[PUBMED]    
3.Weeden SH, Paprosky WG. Minimal 11-year followup of extensively porous-coated stems in femoral revision total hip arthroplasty. J Arthroplasty 2002;17 (4 Suppl 1):134-7.  Back to cited text no. 3
    
4.Zhao X, Yu J, Bu Y, Zheng D. Application of all-coated long stem prosthesis associated with allograft in revision total hip arthroplasty. Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi 2009;23:781-3.  Back to cited text no. 4
[PUBMED]    
5.Kretzer JP, Jakubowitz E, Krachler M, Thomsen M, Heisel C. Metal release and corrosion effects of modular neck total hip arthroplasty. Int Orthop 2009;33:1531-6.  Back to cited text no. 5
[PUBMED]    
6.Moon KH, Kang JS, Lee SH, Jung SR. Revision total hip arthroplasty using an extensively porous coated femoral stem. Clin Orthop Surg 2009;1:105-9.  Back to cited text no. 6
[PUBMED]    
7.Kim YH, Kim JS. Revision hip arthroplasty using strut allografts and fully porous-coated stems. J Arthroplasty 2005;20:454-9.  Back to cited text no. 7
[PUBMED]    
8.Whiteside LA. Major femoral bone loss in revision total hip arthroplasty treated with tapered, porous-coated stems. Clin Orthop Relat Res 2004;429:222-6.  Back to cited text no. 8
[PUBMED]    
9.Engh CA Jr, Ellis TJ, Koralewicz LM, McAuley JP, Engh CA Sr. Extensively porous-coated femoral revision for severe femoral bone loss: Minimum 10-year followup. J Arthroplasty 2002;17:955-60.  Back to cited text no. 9
[PUBMED]    
10.Sporer SM, Paprosky WG. Revision total hip arthroplasty: The limits of fully coated stems. Clin Orthop Relat Res 2003 417:203-9.  Back to cited text no. 10
[PUBMED]    
11.Egan KJ, Di Cesare PE. Intraoperative complications of revision hip arthroplasty using a fully porous-coated straight cobalt-chrome femoral stem. J Arthroplasty 1995;10 Suppl: S45-51.  Back to cited text no. 11
[PUBMED]    

Top
Correspondence Address:
Raju Vaishya
Department of Orthopaedics and Joint Replacement Surgery, Indraprastha Apollo Hospitals, New Delhi
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0019-5413.118211

Rights and Permissions


    Figures

  [Figure 1], [Figure 2], [Figure 3]

This article has been cited by
1 Outcome of a modular head–neck adapter system in revision hip arthroplasty
Maik Hoberg,Christian Konrads,Simon Huber,Stephan Reppenhagen,Matthias Walcher,André Steinert,Thomas Barthel,Maximilian Rudert
Archives of Orthopaedic and Trauma Surgery. 2015; 135(10): 1469
[Pubmed] | [DOI]



 

Top
 
 
 
  Search
 
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Email Alert *
    Add to My List *
* Registration required (free)  
 


 
    Abstract
   Introduction
   Case Reports
   Discussion
    References
    Article Figures
 

 Article Access Statistics
    Viewed1812    
    Printed16    
    Emailed0    
    PDF Downloaded75    
    Comments [Add]    
    Cited by others 1    

Recommend this journal