ORIGINAL ARTICLE  
Year : 2013  |  Volume : 47  |  Issue : 5  |  Page : 443-448
Head size and dislocation rate in primary total hip arthroplasty


Department of Orthopaedics, Civil Hospital and B. J. Medical College, Ahmedabad, Gujarat, India

Click here for correspondence address and email

Date of Web Publication17-Sep-2013
 

   Abstract 

Background: Dislocation after total hip arthroplasty (THA) has a multifactorial etiology with variables such as surgical approach, component orientation and position, type of cup, stem and head size. Review of the literature regarding the relationship of head size and dislocation rate in THA is suggestive that large femoral head size is associated with lower dislocation rate after THA. However, limited data is available as a proof of this hypothesis. The purpose of this study was to determine that the use of large head size would lead to a decreased incidence of dislocations following THA.
Materials and Methods: 317 primary THAs were performed using the posterolateral approach with posterior soft-tissue repair between January 2006 and December 2009. Cases were divided into two groups (A and B). Femoral head diameter size 36 mm was used in 163 THA in group A and 28 mm in 154 THA in group B. Average period of followup being 2 years (6 month to 4 years). Patients were routinely followed at definite intervals and were specifically assessed for dislocation.
Results: One or more dislocations occurred in 11 out of 317 hips with the overall rate of dislocation being 3.47%. Dislocation rate was 0.6% in 36 mm head size and 6.49% with 28 mm head size (P value is 0.0107). Keeping the stem design variable as a constant, the difference in the rate of dislocation between the two groups was again found to be statistically significant for both un-cemented and cemented stem.
Conclusion: Dislocation rate decreased significantly as the size of the head increased in primary THA. However, longer followup is necessary as rate of dislocation or in vivo highly cross linked poly failure or fracture may increase in future affecting the rate of dislocations in primary THA.

Keywords: Dislocation, head size, primary total hip arthroplasty, revision

How to cite this article:
Singh SP, Bhalodiya HP. Head size and dislocation rate in primary total hip arthroplasty. Indian J Orthop 2013;47:443-8

How to cite this URL:
Singh SP, Bhalodiya HP. Head size and dislocation rate in primary total hip arthroplasty. Indian J Orthop [serial online] 2013 [cited 2019 Dec 12];47:443-8. Available from: http://www.ijoonline.com/text.asp?2013/47/5/443/118198

   Introduction Top


The incidence of dislocation after total hip arthroplasty (THA) in several large studies has ranged from 0.4% to 5.8% [1],[2] for primary THA and from 4.8% to 13% for revision total hip arthroplasties. [1],[3],[4] The etiology of dislocation is dictated by patient related factors, surgery-related factors and implant design related variables. [5],[6],[7],[8] Patient related factors include age, sex, height, previous hip surgery, indication for which THA, was carried out neuromuscular impairment and soft-tissue laxity. Surgeon related factors include experience of the surgeon, surgical approach, component orientation, soft-tissue repair, maintenance of leg length and offset; implant design variables include head size, head-neck ratio, head to acetabular ratio, neck geometry and femoral offset. [7] With the availability of newer materials like highly cross linked poly and hard bearings (ceramic, metal on metal), it is now possible to use larger head diameter (36 mm and more) in THA. It has been postulated that the large femoral head diameter reduces the risk of dislocation because of (1) improvement in the head-to-neck ratio, which increases the range of motion of the prosthetic components without prosthetic impingement; (2) a greater amount of translation of the femoral head is required before dislocation occurs as it increases the jump distance; and (3) there may be greater soft-tissue restraints to dislocation as the femoral head is better contained by the surrounding soft tissue envelope. [9],[10],[11],[12] Kelley et al.[11] noted an increased dislocation rate with 22 mm heads as compared to 32 mm heads. They also found that smaller heads, when used with large diameter acetabular components, further increased risk of instability. The use of skirted femoral heads effectively reduces the head-to-neck ratios and contributes to instability. [13] Increased femoral head size had reduced rate of dislocation associated with all operative approaches, but the effect was greatest in association with the posterolateral approach. [8] However, other authors have stated that the size of the femoral head was not related to the dislocation rate. [2],[4],[14] Hedlundh et al. [13],[15] found no significant difference in the dislocation rate between the 22 mm and 32 mm femoral heads. The purpose of this study was to determine whether use of large head size would lead to a decreased incidence of dislocations following THA or not.


   Materials and Methods Top


317 primary THAs performed in 281 patients (245 unilateral, 36 bilateral) with 28 mm or 36 mm diameter femoral head between January 2006 and December 2009 constituted the study material.

Operative and followup records were reviewed for (1) indication for THA (2) type of THA-(cemented, hybrid or uncemented) (3) design of implant (4) head size (5) incidence of dislocation (6) management of dislocation (7) revision surgery for dislocation (8) revision due to any other cause (9) improvement in Harris hip score (HSS). All surgeries were performed by a single surgeon [HPB] using either cemented stem (MS-30, Zimmer, Warsaw, IN, USA)© or uncemented stem (CLS, Zimmer, Warsaw, IN, USA)© . Cups were either ZCA (Zimmer, Warsaw, IN, USA) cemented cup or the Trilogy uncemented cup (Zimmer Warsaw, IN, USA)© . Choice of using cemented or uncemented stem was based on the criteria as shown in [Table 1]. During the study period 36 head size was available only in the un-cemented cup, so all 36 head sizes were un-cemented (n=163), whereas in 28 mm head size cups were both cemented (n=97) and un-cemented (n=57).
Table 1: CLS scoring system (Spotorno criteria)

Click here to view


All patients were operated under either spinal, epidural or the general anesthesia. Posterolateral approach was used for total hip replacement in all cases with enhanced repair of short external rotators and the posterior capsular structures using Ethibond no. 5 anchoring on posterior margins of the greater trochanter. All uncemented cups (n=220) were either press fitted (n=181), or fixation augmented by screws and/or pegs (n=39). All uncemented stems were press fit. Third generation cementing technique was used for cemented stem fixation, employing pressurization, vacuum preparation and the use of a pulsatile lavage. Cup was fixed in 45-50° inclination and 10-15° anteversion. Anteversion of the stem was kept about 10-15°.

All patients were allowed to walk on second postoperative day with a walking frame and by day 4 with the help of a stick. The low molecular weight Heparin was given for 5 days as prophylaxis for deep vein thrombosis. Postoperative limb length, range of motion pain, functional score and total HHS as well as any other observation or complications were assessed and recorded.

Patients were evaluated at 6 months, 1 year and 2 years followup interval. The minimum followup period required for inclusion in this study was kept as 6 months. At each point patients were specifically inquired about any episode of dislocation or any other complications. Dislocation was defined as an event in which the hip required reduction by a clinician.

Cases were divided into two different groups.

  • Group A - 36 mm head size
  • Group B - 28 mm head size.


The demographic distribution of both groups is shown in [Table 2]. The indications of primary THA of both groups are shown in [Figure 1]. In group A, out of 163 hips 64 were MS-30 stem while 99 were CLS stem. In group B, out of 154 hips 94 were MS-30 stem while 60 were CLS stem. Followup was carried out in all cases by either patient's personal visit at the hospital within 2 months of completion of the study phase or by personal visit to patients or by telephonic interview inquiring specifically about the function of hip, any episode of dislocation or any other surgical intervention. All patients (281 patients, 317 hip) were available for the clinicoradiological followup. Radiographs were analyzed for implant position, stem subsidence or migration, osteolysis around stem (zone 1-7 of Gruen) and/or acetabulum (zone 1-3 of DeLee and Charnley).
Figure 1: PI chart showing Indication of THA - difference in group A and group B. (a) Group A, 36 mm head. (b) Group B, 28 mm head

Click here to view
Table 2: Demographic distribution

Click here to view


Statistical analysis

Clinical and radiological data were analyzed using the statistical software epi-info version 3.5.1 released in August 2008 using arithmetic means, standard deviations and confidence interval (CI). P values were evaluated. Fisher test and t test was used to compare the data of the two groups.


   Results Top


Average period of followup was 24 months (range 6-48 months). Out of 317 hips, 173 hips had followup of at least 2 years or more. The average HHS improved overall in both groups. In group A, it improved from 42 preoperatively to 97 at 6 months followup and to 98 at the final followup (2 yrs average, range 6 months to 3 years 7 months). In group B, it improved from 40 preoperatively to 89 at 6 months followup and remained 89 at the final followup (2 years average, range 6 months to 4 years) [Table 3].
Table 3: Comparison between two groups according to HHS with domain

Click here to view


The average range of motion according to Harris hip score improved from 2.16 preoperatively to 4.9 at final followup for group A and from 2.1 preoperatively to 4.1 postoperatively for group B at final followup. The difference in the average range of motion between two groups was not statistically significant.

There were no complications reported in the form of deep vein thrombosis, postoperative infection or periprosthetic fractures, nerve injuries or early loosening during the study period. In our series out of 317 THAs, 11 dislocated, amounting to 3.47% of all within the 1 st year of the study. All 11 cases had posterior dislocation. First episode of dislocation was treated conservatively with closed reduction under anesthesia and examination under image intensifier for a range of stability. All the 11 patients were evaluated radiographically with anteroposterior and cross table lateral views to assess femoral anteversion, acetabular inclination and acetabular version by the method similar to as described by McCollum and Grey and Jolles et al. [16],[17] This was compared for any change in position from the previous X-rays. All patients were given complete bed rest for 6 weeks in abduction. After the rest period, they were rehabilitated gradually to full weight bearing.

In group A, there was 1 dislocation out of a total of 163 (rate of dislocation 0.6%) whereas in group B there were 10 dislocations out of 154 hips (rate of dislocation 6.4%). The P value was 0.0107.

In group A, 1 patient had dislocation, in the first postoperative week and was managed conservatively but had three more episodes of dislocation within 2 months of the first dislocation. He was evaluated by anteroposterior and cross table lateral radiographs of the hip joint after the first episode of dislocation in which his acetabular shell appeared to be in neutral version, but he had 20° of femoral anteversion. Femoral version was calculated by calculating angle formed between the axis of the prosthetic neck and bottom of the radiographic plate representing the posterior plane of the condyles. [16],[17] However, further episodes of dislocation warranted us in depth analysis of acetabular version using a computed tomography (CT) scan. On CT scan his acetabulum component was found to be in 10° of retroversion. This patient underwent revision of the acetabular component [Figure 2].
Figure 2: X-ray left hip joint anteroposterior view showing (a) Dislocation in 36 mm head after total hip replacement due to cup retroversion (b) Revision of uncemented cup with uncemented cup

Click here to view


In group B, 10 patients (6 hips with single episode, 4 hips with multiple episodes of dislocation) had dislocation, all within first 6 weeks of postoperative period except one who presented with dislocation after 3 months with history of recent cerebro vascular accident. Out of 10 cases, 6 were treated conservatively and did well with no further episodes of dislocations or any other complications during the study period. Rest four had more than one episode of dislocation.

One patient had a loss of fixation of the cemented cup after 4 episodes of dislocation in 4 months of postoperative period and was revised from a cemented to an un-cemented cup along with the use of large head. Later his postoperative recovery was good with no further episodes of dislocation. The other 3 patients were further evaluated by CT scan for component position analysis. This group of patients showed lower angles of acetabular anteversion and lower combined anteversion angle values. All the 3 cases needed acetabular revision [Figure 3].
Figure 3: X-ray left hip joint anteroposterior view showing (a) Dislocation in 28 mm head total hip replacement due to lower anteversion of cup with higher abduction angle (b) Revision of cemented cup with cemented cup

Click here to view


The difference in the rate of dislocation among stem variables irrespective of head diameter was 2.51% for CLS stem and 4.43% for MS- 30 stem. The P value was found to be 0.2156. The difference in the rate of dislocation between cemented and un-cemented cups irrespective of the head diameter was 3 (3.09%) out of 94 patients in cemented cup while with un-cemented cup, it was 8 (3.63%) out of 220. The P value was found to be 0.8896 [Table 4].
Table 4: Incidence of dislocation in cemented and cementless stem (Stem variable constant)

Click here to view



   Discussion Top


Along with osteolysis and aseptic loosening, dislocation is one of the most common complications in THA and the most common cause of early revision. Woo and Morrey, Fackler and Poss, Ritter [2],[4],[14] stated that the size of the femoral head was not related to the dislocation rate. Hedlundh et al. [13],[15] saw no significant difference in the dislocation rate between 22 mm and 32 mm femoral heads, but the recurrent dislocation rate increased by 2.3 in the smaller head group. In their study, they used the anterolateral approach with transtrochanteric osteotomy or the posterior approach randomly.

Berry et al. [18] recently reported that in THA, a larger head diameter was associated with a lower long term cumulative risk of dislocation. Malkani et al.[19] assessing 39271 primary THA in Medicare population in USA between 1998 and 2007 which had 2001 number of cases of dislocation; did multivariate cox regression analysis. Other than factors like age, surgeon volume, higher charlson index score (more co morbid condition), they did find a decrease in the rate of dislocation with an increase in use of head size 32 mm or more. During this study period (1998-2007) the proportion of femoral heads, which were size 32 mm or greater, increased from approximately 12 to 77%. Sultan et al. [20] demonstrated that increasing the femoral head diameter from 28 mm to 32 mm essentially increases the arc of motion by 8.2° before impingement and subsequent dislocation. Crowninshield et al. [21] also demonstrated that increasing femoral head diameter using a fixed acetabular abduction angle increased the jump distance or vertical displacement before impingement and dislocation.

Our study represents a series of unselected primary THAs in patients operated for a variety of indications with two different head diameters and two different stem designs. Difference in the rate of dislocation was assessed in terms of head diameter as well as within the stem type. There was no difference between the two groups as far as age, sex or indication of primary THA. All the cases were operated by a single surgeon (HPB) negating the effect of surgeon related factors. Posterolateral approach with enhanced soft-tissue repair was used in all the cases negating the approach related difference among the two groups. Same rehabilitation protocol was followed in both groups. Both groups showed improvement in the HHS. There was no difference in the rate of any complications other than the difference in the rate of dislocation in the two groups.

Our study had two main variables, the difference in head diameter and difference in stem designs. There was no difference in dislocation in between cemented cups and un-cemented cups. However, there was statistically significant difference between the two groups (A and B) in dislocation rates. This difference in the rate of dislocation keeping the stem variable constant was again found to be statistically significant in both un-cemented CLS, as well as for cemented MS 30 stem.

Several studies have already demonstrated the clinical efficacy of larger femoral heads in reducing the dislocation rate. [6],[8],[22],[23],[24],[25] Recently, Bistolfi et al. [26] in their study comparing the risk ratio (RR) of dislocation between the two groups (28 mm vs. 36 mm) found it to be approximately 8, which was a statistically significant. However, the limiting factor in their study was stem design variable which affects neck length and offset. Our study compares, the difference in dislocation keeping the stem design variable as constant. The difference in the rate of dislocation keeping the stem design variable as constant was statistically significant. When stem variable was not considered, the RR was found to be 5.82 (95% CI 0.90-37.85).

One of the major limitations in our study is relatively short duration of followup of the study. Another major drawback of our study was small numbers of patients who were evaluated in each group.

To conclude dislocation rate decreased significantly as the size of the head increased in primary THA. Use of 36 mm diameter head in primary THA also resulted in slightly greater improvement in the range of movements as compared to 28 mm diameter (although statistically not significant). However, longer followup is necessary as rate of dislocation or in vivo highly cross linked poly failure or fracture may increase in future affecting the rate of dislocations in primary THA.

 
   References Top

1.Etienne A, Cupic Z, Charnley J. Postoperative dislocation after Charnley low-friction arthroplasty. Clin Orthop Relat Res 1978;132:19-23.  Back to cited text no. 1
[PUBMED]    
2.Woo RY, Morrey BF. Dislocations after total hip arthroplasty. J Bone Joint Surg Am 1982;64:1295-306.  Back to cited text no. 2
[PUBMED]    
3.Daly PJ, Morrey BF. Operative correction of an unstable total hip arthroplasty. J Bone Joint Surg Am 1992;74:1334-43.  Back to cited text no. 3
[PUBMED]    
4.Fackler CD, Poss R. Dislocation in total hip arthroplasties. Clin Orthop Relat Res 1980;151:169-78.  Back to cited text no. 4
[PUBMED]    
5.Meek RM, Allan DB, McPhillips G, Kerr L, Howie CR. Late dislocation after total hip arthroplasty. Clin Med Res 2008;6:17-23.  Back to cited text no. 5
[PUBMED]    
6.Meek RM, Allan DB, McPhillips G, Kerr L, Howie CR. Epidemiology of dislocation after total hip arthroplasty. Clin Orthop Relat Res 2006;447:9-18.  Back to cited text no. 6
[PUBMED]    
7.Sanchez-Sotelo J, Berry DJ. Epidemiology of instability after total hip replacement. Orthop Clin North Am 2001;32:543-52.  Back to cited text no. 7
[PUBMED]    
8.Berry DJ, von Knoch M, Schleck CD, Harmsen WS. Effect of femoral head diameter and operative approach on risk of dislocation after primary total hip arthroplasty. J Bone Joint Surg Am 2005;87:2456-63.  Back to cited text no. 8
[PUBMED]    
9.Amstutz HC, Lodwig RM, Schurman DJ, Hodgson AG. Range of motion studies for total hip replacements. A comparative study with a new experimental apparatus. Clin Orthop Relat Res 1975;111:124-30.  Back to cited text no. 9
[PUBMED]    
10.Eftekhar NS. Dislocation and instability complicating low friction arthroplasty of the hip joint. Clin Orthop Relat Res 1976;121:120-5.  Back to cited text no. 10
[PUBMED]    
11.Kelley SS, Lachiewicz PF, Hickman JM, Paterno SM. Relationship of femoral head and acetabular size to the prevalence of dislocation. Clin Orthop Relat Res 1998;355:163-70.  Back to cited text no. 11
[PUBMED]    
12.Morrey BF. Instability after total hip arthroplasty. Orthop Clin North Am 1992;23:237-48.  Back to cited text no. 12
[PUBMED]    
13.Hedlundh U, Ahnfelt L, Hybbinette CH, Wallinder L, Weckstrom J, Fredin H. Dislocations and the femoral head size in primary hip arthroplasty. Clin Orthop 1996;333:226-33.  Back to cited text no. 13
    
14.Ritter MA. A treatment plan for the dislocated total hip arthroplasty. Clin Orthop Relat Res 1980;155:153-5.  Back to cited text no. 14
    
15.15. Hedlundh U, Ahnfelt L, Hybbinette CH, Weckstrom J, Fredin H. Surgical experience related to dislocations after total hip arthroplasty. J Bone Joint Surg Br 1996;78:2 06-9.  Back to cited text no. 15
    
16.McCollum DE, Gray WJ. Dislocation after total hip arthroplasty. Causes and prevention. Clin Orthop Relat Res 1990;261:159-70.  Back to cited text no. 16
[PUBMED]    
17.Jolles BM, Zangger P, Leyvraz PF. Factors predisposing to dislocation after primary total hip arthroplasty: A multivariate analysis. J Arthroplasty 2002;17:282-8.  Back to cited text no. 17
[PUBMED]    
18.Berry DJ, von Knoch M, Schleck CD, Harmsen WS. The cumulative long term risk of dislocation after primary Charnley total hip arthroplasty. J Bone Joint Surg Am 2004;86-A: 9-14.  Back to cited text no. 18
[PUBMED]    
19.Malkani AL, Ong KL, Lau E, Kurtz SM, Justice BJ, Manley MT. Early- and late-term dislocation risk after primary hip arthroplasty in the Medicare population. J Arthroplasty 2010;25 Suppl 1:21-5.  Back to cited text no. 19
    
20.Sultan PG, Tan V, Lai M, Garino JP. Independent contribution of elevated-rim acetabular liner and femoral head size to the stability of total hip implants. J Arthroplasty 2002;17:289-92.  Back to cited text no. 20
[PUBMED]    
21.Crowninshield RD, Maloney WJ, Wentz DH, Humphrey SM, Blanchard CR. Biomechanics of large femoral heads: What they do and don›t do. Clin Orthop Relat Res 2004;429:102-7.  Back to cited text no. 21
[PUBMED]    
22.Burroughs BR, Hallstrom B, Golladay GJ, Hoeffel D, Harris WH. Range of motion and stability in total hip arthroplasty with 28-, 32-, 38- and 44-mm femoral head sizes. J Arthroplasty 2005;20:11-9.  Back to cited text no. 22
[PUBMED]    
23.Byström S, Espehaug B, Furnes O, Havelin LI, Norwegian Arthroplasty Register. Femoral head size is a risk factor for total hip luxation: A study of 42,987 primary hip arthroplasties from the Norwegian Arthroplasty Register. Acta Orthop Scand 2003;74:514-24.  Back to cited text no. 23
    
24.Clarke IC, Gustafson A, Jung H, Fujisawa A. Hip-simulator ranking of polyethylene wear: Comparisons between ceramic heads of different sizes. Acta Orthop Scand 1996;67:128-32.  Back to cited text no. 24
[PUBMED]    
25.Kluess D, Martin H, Mittelmeier W, Schmitz KP, Bader R. Influence of femoral head size on impingement, dislocation and stress distribution in total hip replacement. Med Eng Phys 2007;29:465-71.  Back to cited text no. 25
[PUBMED]    
26.Bistolfi A, Crova M, Rosso F, Titolo P, Ventura S, Massazza G. Dislocation rate after hip arthroplasty within the first postoperative year: 36 mm versus 28 mm femoral heads. Hip Int 2011;21:559-64.  Back to cited text no. 26
[PUBMED]    

Top
Correspondence Address:
Somesh P Singh
Department of Orthopedics, GMERS Medical College and Hospital, Sola, Ahmedabad - 380 060, Gujarat
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0019-5413.118198

Rights and Permissions


    Figures

  [Figure 1], [Figure 2], [Figure 3]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4]

This article has been cited by
1 Uniaxial Compaction-Based Manufacturing Strategy and 3D Microstructural Evaluation of Near-Net-Shaped ZrO2-Toughened Al2O3Acetabular Socket?
Debasish Sarkar,Bhimavarapu Sambi Reddy,Sourav Mandal,Mamilla RaviSankar,Bikramjit Basu
Advanced Engineering Materials. 2016; 18(9): 1634
[Pubmed] | [DOI]
2 Is There a Benefit to Head Size Greater Than 36 mm in Total Hip Arthroplasty?
Bryan D. Haughom,Darren R. Plummer,Mario Moric,Craig J. Della Valle
The Journal of Arthroplasty. 2016; 31(1): 152
[Pubmed] | [DOI]
3 The Divergence of Wear Propagation and Stress at Steep Acetabular Cup Positions Using Ceramic Heads and Sequentially Cross-Linked Polyethylene Liners
Carmen Zietz,Christian Fabry,Felix Baum,Rainer Bader,Daniel Kluess
The Journal of Arthroplasty. 2015;
[Pubmed] | [DOI]
4 Dislocation Rates Following Primary Total Hip Arthroplasty Have Plateaued in the Medicare Population
Akshay Goel,Edmund Lau,Kevin L. Ong,Daniel J. Berry,Arthur L. Malkani
The Journal of Arthroplasty. 2014;
[Pubmed] | [DOI]



 

Top
 
 
 
  Search
 
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Email Alert *
    Add to My List *
* Registration required (free)  
 


 
    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
    References
    Article Figures
    Article Tables
 

 Article Access Statistics
    Viewed7255    
    Printed52    
    Emailed3    
    PDF Downloaded311    
    Comments [Add]    
    Cited by others 4    

Recommend this journal