CASE REPORT  
Year : 2013  |  Volume : 47  |  Issue : 1  |  Page : 98-100
A vascular complication in computer navigated total knee arthroplasty


Department of Orthopaedic Surgery, Northwick Park Hospital, Watford Road, Harrow, London, United Kingdom

Click here for correspondence address and email

Date of Web Publication4-Feb-2013
 

   Abstract 

The use of computer navigation has the potential to improve implant position in total knee arthroplasty (TKA), but pin fixation of reference arrays introduces an additional potential source of complications. We report a case of vascular injury related to the insertion of a femoral pin during navigated TKA.

Keywords: Navigated knee arthroplasty, total knee arthroplasty, vascular complications

How to cite this article:
Gulhane S, Holloway I, Bartlett M. A vascular complication in computer navigated total knee arthroplasty . Indian J Orthop 2013;47:98-100

How to cite this URL:
Gulhane S, Holloway I, Bartlett M. A vascular complication in computer navigated total knee arthroplasty . Indian J Orthop [serial online] 2013 [cited 2019 Dec 10];47:98-100. Available from: http://www.ijoonline.com/text.asp?2013/47/1/98/106927

   Introduction Top


Computer-assisted surgery has been popularized in orthopedic surgery in the last 10 years. Whilst a number of recent randomized control trials have shown some benefit in terms of implant position over conventional jig-based referencing, [1] improved survivorship has not yet been demonstrated.

Many navigation systems require the use of reference arrays, which are fixed to the tibia and femur with pins. Reported complications include pin breakage, [2] superficial wound infection, [3],[4] prolonged operation time, [5],[6],[7],[8] and prolonged tourniquet time. [8]


   Case Report Top


A 58-year-old man with a diagnosis of varus osteoarthritis (OA) of the right knee underwent navigated total knee arthroplasty (TKA) using standard technique.

Our current practice is to use computer navigation for all primary TKA. We use a passive reflector-based system (Brainlab, Feldkirchen, Germany) with fixation of the femoral reference array using two 3-mm-diameter threaded pins. These are inserted anteriorly through stab incisions, with the knee in flexion, as proximally as the tourniquet will allow. Tibial pins are inserted in the mid diaphysis into the subcutaneous surface of the tibia [Figure 1]. All pins are bicortical in order to maintain secure fixation for the duration of surgery.
Figure 1: Peroperative photographs (a) Insertion of tibial and femoral reference arrays (b) The styloid used to map out the tibia and femur

Click here to view


Computer registration is performed with digitalization of the hip and ankle centers and specific landmarks of the knee. A size 6 generation II cemented femoral implant and a size 6 generationII cemented tibial component were implanted with an 11-mm polyethylene insert. There was no obvious complication during surgery.

The patient had some pain and thigh swelling postoperatively, but this was considered to be in keeping with recent surgery. He was discharged on the third postoperative day, having satisfied physiotherapy and occupational therapy goals.

He reported 3 days later with increased right thigh pain, swelling, and a limited range of movement. He was given analgesia and placed onto a continuous passive motion machine. Deep vein thrombosis and fracture were excluded on ultrasound scan and plain radiographs. An ultrasound guided diagnosis of a quadriceps hematoma at the site of the femoral pins was made. This was confirmed by computerized topography (CT) with intravenous contrast [Figure 2], which showed a large quadriceps hematoma and active bleeding into the femoral canal from a branch of the superficial femoral artery as it entered linea aspera at the site of pin insertion.
Figure 2: Two CT scan images showing active bleeding into the femoral canal

Click here to view


On day 10, the hematoma was evacuated and two 14-mm unicortical screws were inserted into the femur at the site of the bleed to tamponade the bleed. Unicortical screws were used to reduce the risk of further arterial injury with bicortical screws and the CT scan also showed unicortical injury.

At 2 weeks post presentation, he underwent a manipulation under anesthesia for ongoing stiffness. He then had a flexion of 0°-90°, but also had some persistent swelling and was prescribed a short course of diazepam. He showed continued improvement since the second operation and had no further complications at the final followup 6 months later [Figure 3].
Figure 3: Final followup X-ray showing insertion of two unicortical screws

Click here to view



   Discussion Top


Computer navigated TKA has the potential advantage of improved implant position, which is known to reduce the rate of aseptic loosening. There have been a number of studies reporting complications of computer navigated TKA among these. Böhling et al., [9] Chin et al., [10] and Bäthis et al. [11] demonstrated longer tourniquet time with computer navigation.

Kalairajah et al. [12] reported a longer tourniquet time as well as a statistically significant decrease in blood drainage. Chauhan et al. [13] and Jenny et al. [4] found that computerized navigation was associated with a longer operating time.

Stöckl et al. [14] reported one case of poor tibial tracker stability, related to osteoporotic bone, resulting in abandonment of computer navigation, from 64 patients. Victor et al. [2] found pin breakage in 3 patients from a population of 100. Hernañdez-Vaquero [15] reported four cases of failed pin fixation (three iliac crest and one femur) requiring abandonment of computer navigation from a total of 112 patients.

Our report is the only case of arterial injury in the literature related to navigation pin usage. It highlights an important complication of computer navigated TKA. If the patient develops an unexpected postoperative swelling with increasing pain this diagnosis needs to be taken into consideration.

 
   References Top

1.Bauwens K, Matthes G, Wich M, Gebhard F, Hanson B, Ekkernkamp A, et al. Navigated total knee replacement. A meta-analysis. J Bone Joint Surg Am 2007;89:261-9.  Back to cited text no. 1
[PUBMED]    
2.Victor J, Hoste D. Image-based computer-assisted total knee arthroplasty leads to lower variability in coronal alignment. Clin Orthop Relat Res 2004;428:131-9.  Back to cited text no. 2
[PUBMED]    
3.Cossey AJ, Spriggins AJ. The use of computer-assisted surgical navigation to prevent malalignment in unicompartmental knee arthroplasty. J Arthroplasty 2005;20:29-34.  Back to cited text no. 3
[PUBMED]    
4.Jenny JY, Clemens U, Kohler S, Kiefer H, Konermann W, Miehlke RK. Consistency of implantation of a total knee arthroplasty with a non-image-based navigation system: A case-control study of 235 cases compared with 235 conventionally implanted prostheses. J Arthroplasty 2005;20:832.  Back to cited text no. 4
[PUBMED]    
5.Perlick L, Bäthis H, Tingart M, Perlick C, Luring C, Grifka J. Minimally invasive unicompartmental knee replacement with a nonimage-based navigation system. Int Orthop 2004;28:193-7.  Back to cited text no. 5
    
6.Hart R, Janeèek M, Chaker A, Buèek P. Total knee arthroplasty implanted with and without kinematic navigation. Int Orthop 2003;27:366-9.  Back to cited text no. 6
    
7.Zorman D, Etuin P, Jennart H, Scipioni D, Devos S. Computer-assisted total knee arthroplasty: Comparative results in a preliminary series of 72 cases. Acta Orthop Belg 2005;71:696-702.  Back to cited text no. 7
[PUBMED]    
8.Decking R, Markmann Y, Fuchs J, Puhl W, Scharf HP. Leg axis after computer-navigated total knee arthroplasty: A prospective randomized trial comparing computer-navigated and manual implantation. J Arthroplasty 2005;20:282-8.  Back to cited text no. 8
[PUBMED]    
9.Böhling U, Schamberger H, Grittner U, Scholz J. Computerised and technical navigation in total knee-arthroplasty. J Orthopaed Traumatol 2005;6:69-75.  Back to cited text no. 9
    
10.Chin PL, Yang KY, Yeo SJ, Lo NN. Randomized control trial comparing radiographic total knee arthroplasty implant placement using computer navigation versus conventional technique. J Arthroplasty 2005;20:618-26.  Back to cited text no. 10
    
11.Bäthis H, Perlick L, Tingart M, Lüring C, Zurakowski D, Grifka J. Alignment in total knee arthroplasty. A comparison of computer-assisted surgery with the conventional technique. J Bone Joint Surg Br 2004;86:682-7.  Back to cited text no. 11
    
12.Kalairajah Y, Simpson D, Cossey AJ, Verrall GM, Spriggins AJ. Blood loss after total knee replacement: Effects of computer assisted surgery. J Bone Joint Surg Br 2005;87:1480-2.  Back to cited text no. 12
    
13.Chauhan SK, Scott RG, Breidahl W, Beaver RJ. Computer-assisted knee arthroplasty versus a conventional jig-based technique. A randomised, prospective trial. J Bone Joint Surg Br 2004;86:372-7.  Back to cited text no. 13
    
14.Stöckl B, Nogler M, Rosiek R, Fischer M, Krismer M, Kessler O. Navigation improves accuracy of rotational alignment in total knee arthroplasty. Clin Orthop Relat Res 2004;426:180-6.  Back to cited text no. 14
    
15.Hernandez-Vaquero D, Suuarez-vazquez A. Complications of fixed infrared emitters in computer-assisted total knee arthroplasties. BMC Musculoskelet Disord 2007;8:71.  Back to cited text no. 15
    

Top
Correspondence Address:
Sandesh Gulhane
22 Temple Gardens, Goldersgreen, London, NW11 0LL
United Kingdom
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0019-5413.106927

Rights and Permissions


    Figures

  [Figure 1], [Figure 2], [Figure 3]

This article has been cited by
1 Is determination of the hip rotation center during computer-assisted surgery influenced by the pinless femoral sensor attachment method? An in vitro preliminary study comparing osseous pin fixation
E. de Thomasson,S. Gouzy,F. Cloerec,L. Geais
Orthopaedics & Traumatology: Surgery & Research. 2016; 102(3): 375
[Pubmed] | [DOI]
2 Computer-assisted navigation is beneficial both in primary and revision surgery with modular rotating-hinge knee arthroplasty
Björn G. Ochs,Anna J. Schreiner,Peter M. de Zwart,Ulrich Stöckle,Christoph Emanuel Gonser
Knee Surgery, Sports Traumatology, Arthroscopy. 2014;
[Pubmed] | [DOI]
3 Clinical application of minimally invasive total knee arthroplasty
Gao, Y.-J., Wang, W.-L.
Source of the Document Chinese Journal of Tissue Engineering Research. 2013; 17(39): 6985-6990
[Pubmed]



 

Top
 
 
 
  Search
 
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Email Alert *
    Add to My List *
* Registration required (free)  
 


 
    Abstract
   Introduction
   Case Report
   Discussion
    References
    Article Figures
 

 Article Access Statistics
    Viewed1730    
    Printed37    
    Emailed3    
    PDF Downloaded62    
    Comments [Add]    
    Cited by others 3    

Recommend this journal