Results of uncemented distal locked prosthesis in revision hip arthroplasty with proximal femoral bone loss: A retrospective study
|
|
Sanjay Agarwala1, Hari Ram Jhunjhunwala2, Jawahir A Pachore2, Siddharth B Joglekar3, Krishna Kiran Eachempati4
1 Department of Orthopaedics and Traumatology and Chief of Surgery, P D Hinduja National Hospital and Medical Research Centre, Mumbai, India 2 Department of Orthopaedic Surgery, Bombay Hospital and Medical Research Centre, Mumbai, India 3 University of Minnesota, Minneapolis, MN, USA 4 Consultant Orthopedic Surgeon, Krishna Institute of Medical Sciences, Secunderabad, India
Click here for correspondence address and email
Date of Web Publication | 4-Feb-2013 |
|
|
 |
|
Abstract | | |
Introduction: Management of bone loss is a challenge in revision total hip arthroplasty (THA). A retrospective review was performed to study the use of uncemented distal locked prosthesis in cases with proximal femoral bone loss. Materials and Methods: Uncemented stems with distal interlocking were used in 65 hips during revision THAs with 38 hips having Paprosky IIIB/IV defects between January 1998 and February 2004. There were 48 males and 17 females in the study with an average age of 53 years (range 30-80 years). Radiographic and clinical outcome evaluation using the Harris hip score (HHS) were performed. Results: An improvement in HHS (mean: 33 points) was observed at final followup (mean: 9 years). Regeneration of proximal bone stock was observed without signs of loosening or subsidence and none of the stems were revised. Three patients developed recurrent dislocation while one had a stem subsidence of 1cm following removal of interlocking bolts. Conclusion: Uncemented distal locked prosthesis provide adequate stability in revision THA, aiding the reconstruction of bony deficiencies while avoiding the disadvantages of fully porous or cemented implants. Keywords: Distal locked prosthesis, femoral bone loss, revision hip arthroplasty, uncemented revision total hip arthroplasty
How to cite this article: Agarwala S, Jhunjhunwala HR, Pachore JA, Joglekar SB, Eachempati KK. Results of uncemented distal locked prosthesis in revision hip arthroplasty with proximal femoral bone loss: A retrospective study
. Indian J Orthop 2013;47:83-6 |
How to cite this URL: Agarwala S, Jhunjhunwala HR, Pachore JA, Joglekar SB, Eachempati KK. Results of uncemented distal locked prosthesis in revision hip arthroplasty with proximal femoral bone loss: A retrospective study
. Indian J Orthop [serial online] 2013 [cited 2019 Dec 7];47:83-6. Available from: http://www.ijoonline.com/text.asp?2013/47/1/83/106918 |
Introduction | |  |
The commonest mechanism of failure of THA is aseptic loosening apart from other causes like periprosthetic fracture, stem fracture, infection, and recurrent dislocation. [1] Femoral revision is complicated by loss of the proximal femoral bone. The goal of revision is to obtain fixation of the revision implant and to reconstruct the proximal femoral bony deficiency.
The grade and type of the bony defects in the proximal femur are important determinants in preoperative planning and selection of implants. A variety of options regarding implant choice and method of fixation are available for revision THA. [2] The ideal surgical technique would aim to provide adequate stability while preserving and reconstructing deficient bone stock. It is also important that the implant helps to prevent further loss of bone.
A distal locked prosthesis has been shown to be successful in the revisions using the "reversed fixation principle." The distal locking provides initial stability necessary for the natural process of bone healing and osseointegration. After the reconstruction of proximal bone loss, the interlocking bolts are removed, thus transferring the load delivery back to the proximal femur, which is more physiological. [3]
We present our experience following use of a proximally coated distal locked prosthesis in revision hip arthroplasty at an average of 9 years postoperatively.
Materials and Methods | |  |
Between January 1998 and February 2004, 82 revision THAs were performed using the BiCONTACT revision stem (B.Braun-Aesculap, Tuttlingen, Germany) by four senior authors (SA, HRJ, JAP, SBJ) at two medical centers. All four surgeons were not always a part of the same operating team. The patient's data were recorded prior to surgery and at every followup. Of these, six patients died of unrelated causes and ten patients were lost to followup, one patient who underwent a 2-stage revision for infection was excluded. The remaining 65 patients were evaluated in February 2012 from the subject of the present study.
Patients who underwent revision of the femoral component with Bicontact interlocking stem for aseptic loosening or periprosthetic fracture of femur and were available for followup between January and February 2012 were included in the study. Patients who underwent surgery for infected revision, patients who were lost to followup or died were excluded. The implant used was an uncemented tapered proximally plasma sprayed stem with a facility for two distal interlocking bolts, of which the most distal slot allows dynamization. All the patients were informed preoperatively regarding the type of implant, which they would receive. A written informed consent was obtained. They were also informed that the data regarding their treatment would be submitted for publication and an approval for data collection was obtained from the institutional review board. All patients were reviewed annually and the final followup was in 2012. A total of 66 patients attended followup; 65 of these were included in the study and one patient who had a 2-stage revision for sepsis was excluded.
There were 48 males and 17 females in the study with an average age of 53 years (range 30-80 years). Forty four cases with both component loosening had both femoral and acetabular revision. Fifty-six cases had aseptic loosening, whereas nine cases had a periprosthetic fracture. Six cases had a Vancouver type B3 fracture and three cases had a type B2 fracture. [4]
In 40 cases, the primary component was cemented and uncemented in 25 cases. Patients with isolated femoral component loosening (n = 12) or periprosthetic femur fracture (n = 9) had femoral revision only while those with both femoral and acetabular loosening (n = 44) had both femoral and acetabular revision. In four cases, well-fixed acetabulum had to be removed to change version and address intraoperative instability. The proximal femoral bone loss was graded using the Paprosky classification. [5] There were 10 cases with grade II Paprosky defects, 18 cases with grade IIIA defects, 26 cases with grade IIIB defects [Figure 1], and 12 cases with grade IV defects [Figure 2]. | Figure 1: Preoperative X-ray (anteroposterior view) of left hip (a) showing grade III b Paprosky defect managed with isolated femoral revision 7 years followup X-ray (b) shows no subsidence.
Click here to view |
 | Figure 2: Preoperative anteroposterior view of right hip (a) shows grade IV Paprosky defect managed with isolated femoral revision immediate postoperative (b) and 9 years followup (c) X-Rays after bolt removal showing good proximal bone healing and no subsidence
Click here to view |
Operative procedure
The lateral position was used in all the patients. Fifty five hips were operated by posterior approach while 11 by anterolateral approach.
In 13 cases, the implant along with the entire cement mantle could be extracted from the proximal femur without any need for an osteotomy; 25 cases required a transfemoral approach for implant removal. In 27 cases, a ventral cortical window was sufficient. [6] Prophylactic cerclage wiring just distal to the osteotomy was used to prevent splintering at the time of the insertion of the prosthesis. The distal femur was gently reamed with a hand reamer to preserve bone. A BiCONTACT stem (B.Braun-Aesculap, Tuttlingen, Germany) of an adequate length was then introduced bridging the defect zone by at least 4-6 cm and cerclage closure of the osteotomy site was performed. The distal interlocking was performed depending upon the grade of proximal femoral deficiency and quality of distal fixation. No interlocking was performed in 5 cases. Interlocking was optional in case of Paprosky Type II and IIIA defects and mandatory in case of Type IIIB and IV defects. Two interlocking bolts were used in 43 cases, whereas only the most distal interlocking slot was used in 18 cases. Image guidance or the jig provided by the manufacturer was used for interlocking and one or two bolts were used according to the preference of the operating surgeon. Morselized allografts were packed around the implant at the time of introduction and slurry grafts (product of acetabular reaming) were used around the osteotomy. Interlocking bolts were removed at a mean of 14 months (range 6-24 months) postoperatively.
The average hospitalization time of the patients was 16 days. The patients with a transfemoral osteotomy (n = 25) were allowed only nonweight bearing ambulation using a walker for the first 6 weeks followed by toe touch ambulation for a further 6 weeks. In cases where a cortical window was made (n = 27), toe touch ambulation with the help of a walker was started as soon as the patients were comfortable. An abduction brace was used for ambulation in the first 2 weeks following surgery in 41 cases. An abduction pillow was used when in bed during the same period.
Followups were at 6 weekly intervals for the first 6 months and then annually, with the latest followup between May and June 2012. Radiological and clinical evaluation including Harris hip score was performed at every visit. [6] The radiographs were evaluated for evidence of fixation and signs of instability according to the criteria of Engh et al. apart from remodeling and reconstruction of the proximal femoral bone stock. [7]
The criteria of Gross et al. were used to define a successful result in our series. [8] These criteria include a postoperative increase in the Harris hip score of more than 20 points, a radiologically stable implant, and no need for further femoral reconstructive surgery. Additionally, the classification of a successful result included signs of regeneration of the proximal femur and absence of stress shielding.
Results | |  |
Average followup was 9 years (range 8-14 years) with no re-revision being required at the time of final followup for any stem. There was an improvement in Harris hip scores from a mean value of 41 to 74 points.
The radiological evaluation did not reveal proximal stress shielding in any case. However, there was evidence of distal cortical hypertrophy in cases where the interlocking bolts were not removed. None of the cases showed any evidence of extensive reactive lines or a pedestal sign. Spot welds were seen adjacent to the proximal plasma sprayed surface of the stems in 43 cases. Average subsidence of approximately 10 mm was seen in stems without interlocking (range 0-12 mm). There was one subsidence in cases where interlocking was performed after secondary removal of the bolts. Proximal bone remodeling and restoration was visible in all the cases.
Complications
Three cases developed recurrent dislocation, which was managed by cup revision in two and use of 28 mm liner instead of 22 mm and +6 mm head in the third case. One patient developed about 1 cm subsidence after removal of the interlocking bolts at 2 years. This patient had a revision for periprosthetic fracture about an isoelastic stem, proximally tibial allograft from TKR bone cut was used to reconstruct a proximal femoral deficiency. The fracture had healed successfully. However, the hip remained stable and the patient continued to walk with an abductor lurch. None of the cases developed deep infection, periprosthetic fracture, or aseptic loosening.
Discussion | |  |
The poor proximal bone stock in revision hip surgery requires implants that give fixation distal to the bony defects. [2] Implants which are extensively porous coated lead to further stress shielding of the proximal femur due to the nonphysiological load transfer. [9],[10]
Cemented fixation is not ideal in such cases as there is not enough cancellous bone stock in the proximal femur to secure an adequate micro-interlock of the cement to the bone. [11] The stems, which are proximally coated, lack initial stability and fail by subsidence. [12]
The BiCONTACT system (B.Braun-Aesculap, Tuttlingen, Germany) using the "reversed fixation" principle provides primary stability with the option to transfer load delivery back to the proximal femur secondarily. The initial torsional and axial stability of this system favors healing and remodeling of the proximal femur. The removal of interlocking bolts contributes to physiological load transfer to the proximal metaphyseal area of the femur and prevents further loss of bone stock due to stress shielding. [3],[13]
An ideal study would include a prospective randomized controlled trial to evaluate the efficacy of distal locking in revision THA. A retrospective study has inherent limitations and bias. However, it does provide useful information and directions for further research. Our study indicated excellent results with use of distal locked implants. None of the cases were revised and only one demonstrated loosening. Subsidence was noted when distal locking was not performed in cases with low-grade bone loss. This underlines the need for primary stability with distal locking in these implants. Routine removal of interlocking bolts was performed at two years to transfer the load proximally. This correlates well with various authors who have presented their results with such prosthesis in the past.
Mahomed et al. reported that distal interlocking of a press fit femoral prosthesis increased the torsional and axial stability which led to improvement in biological fixation. [14] Kim et al. reported the results of 68 femoral revisions done with a cementless proximally porous-coated stem with distal interlocking screws (BiCONTACT: B.Braun-Aesculap, Tuttlingen, Germany). At final followup at a mean of 40 months, there was no subsidence or loosening in their series. [15] Volkmann et al. reported a calculated survival rate of 85.3% with an increase in the Harris hip scores from 42 to 75 points in their series of 109 cases with a mean followup of 5.25 years. [3] Eingartner et al. reported similar encouraging results. [13] Learmonth et al. have reported results of revision hip arthroplasty using the BiCONTACT system (B.Braun-Aesculap, Tuttlingen, Germany) in 22 cases with Vancouver type B2 periprosthetic fractures. [16] We had six cases with Vancouver type B3 fractures and three cases with Vancouver type B2 fracture in our series. We have experienced a successful result with these cases in our series without complications. Sotereanos et al. [17] have also demonstrated the effectiveness of distal locking with a 94% success rate in their series. However, a custom-built prosthesis was used in their series. Furthermore, the stems were extensively porous coated and therefore lack the advantage of the BiCONTACT system in preventing further proximal bone loss. The BiCONTACT system, using the "reversed fixation" principle, provides primary stability with the option to transfer load delivery back to the proximal femur secondarily. This is a biological method of revision as claimed by Volkmann et al. [3],[13] In our study, there was no stress shielding visible even in cases without removal of the interlocking bolts in spite of the cortical hypertrophy seen around the distal end of the prosthesis. The absence of subsidence after secondary interlock bolt removal is evidence of osseointegration leading to secondary stability.
To conclude, a distal locked system is a useful implant system while dealing with revision hip cases with high-grade proximal femoral defects. It provides adequate stability in reconstruction of proximal femoral bone loss. Further prospective studies will help to determine whether these implants are superior to the currently available implants used in revision hip surgery.
References | |  |
1. | Madey SM, Callaghan JJ, Olejniczak JP, Goetz DD, Johnston RC. Charnley total hip arthroplasty with use of improved techniques of cementing. The results after a minimum of fifteen years of followup. J Bone Joint Surg Am 1997;79:53-64.  [PUBMED] |
2. | Maurer SG, Baitner AC, Di Cesare PE. Reconstruction of the failed femoral component and proximal femoral bone loss in revision hip surgery. J Am Acad Orthop Surg 2000;8:354-63.  [PUBMED] |
3. | Volkmann R, Bretschneider C, Eingartner C, Weller S. Revision arthroplasty-femoral aspect: The concept to solve high grade defects. Int Orthop 2003;27 Suppl 1:S24-8.  [PUBMED] |
4. | Masri BA, Meek RM, Duncan CP. Periprosthetic fractures evaluation and treatment. Clin Orthop Relat Res 2004;420:80-95.  [PUBMED] |
5. | Valle CJ, Paprosky WG. Classification and an algorithmic approach to the reconstruction of femoral deficiency in revision total hip arthroplasty. J Bone Joint Surg Am 2003;85-A Suppl 4:1-6.  [PUBMED] |
6. | Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures: Treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am 1969;51:737-55.  [PUBMED] |
7. | Engh CA, Massin P, Suthers KE. Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. Clin Orthop Relat Res 1990;257:107-28.  [PUBMED] |
8. | Gross AE, Hutchison CR, Alexeeff M, Mahomed N, Leitch K, Morsi E. Proximal femoral allografts for reconstruction of bone stock in revision arthroplasty of the hip. Clin Orthop Relat Res 1995;319:151-8.  [PUBMED] |
9. | Krishnamurthy AB, MacDonald SJ, Paprosky WG. 5- to 13-year followup study on cementless femoral components in revision surgery. J Arthroplasty 1997;12:839-47.  [PUBMED] |
10. | Lawrence JM, Engh CA, Macalino GE, Lauro GR. Outcome of revision hip arthroplasty done without cement. J Bone Joint Surg Am 1994;76:965-73.  [PUBMED] |
11. | Dohmae Y, Bechtold JE, Sherman RE, Puno RM, Gustilo RB. Reduction in cement-bone interface shear strength between primary and revision arthroplasty. Clin Orthop Relat Res 1988;236:214-20.  [PUBMED] |
12. | Peters CL, Rivero DP, Kull LR, Jacobs JJ, Rosenberg AG, Galante JO. Revision total hip arthroplasty without cement: Subsidence of proximally porous-coated femoral components. J Bone Joint Surg Am 1995;77:1217-26.  [PUBMED] |
13. | Eingartner C, Volkmann R, Putz M, Weller S. Uncemented revision stem for biological osteosynthesis in periprosthetic femoral fractures. Int Orthop 1997;21:25-9.  |
14. | Mahomed N, Schatzker J, Hearn T. Biomechanical analysis of a distally interlocked press-fit femoral total hip prosthesis. J Arthroplasty 1993;8:129-32.  [PUBMED] |
15. | Kim YM, Kim HJ, Song WS, Yoo JJ. Experiences with the BiCONTACT revision stems with distal interlocking. J Arthroplasty 2004;19:27-34.  [PUBMED] |
16. | Learmonth ID. The management of periprosthetic fractures around the femoral stem. J Bone Joint Surg Br 2004;86:13-9.  [PUBMED] |
17. | Sotereanos N, Sewecke J, Raukar GJ, DeMeo PJ, Bargiotas K, Wohlrab D. Revision total hip arthroplasty with a custom cementless stem with distal cross-locking screws. Early results in femora with large proximal segmental deficiencies. J Bone Joint Surg Am 2006;88:1079-84.  [PUBMED] |

Correspondence Address: Jawahir A Pachore Director Hip Surgery, Shalby Hospital, Karnavati Club, Sabarmati-Gandhi Nagar Highway, Ahmedabad - 380 015, Gujarat India
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/0019-5413.106918

[Figure 1], [Figure 2] |
|
This article has been cited by | 1 |
Are Range of Motion Measurements Needed When Calculating the Harris Hip Score? |
|
| Paul K. Edwards,Robin M. Queen,Robert J. Butler,Michael P. Bolognesi,C. Lowry Barnes | | The Journal of Arthroplasty. 2016; 31(4): 815 | | [Pubmed] | [DOI] | | 2 |
Mid-Term Outcomes and Complications with Cementless Distal Locking Hip Revision Stem with Hydroxyapatite Coating for Proximal Bone Defects and Fractures |
|
| Lluis Carrera,Sleiman Haddad,Joan Minguell,Carles Amat,Pablo S. Corona | | The Journal of Arthroplasty. 2015; | | [Pubmed] | [DOI] | | 3 |
Indian Journal of orthopaedics: Journey continues |
|
| Jain, A.K. and Kumar, S. | | Indian Journal of Orthopaedics. 2013; 47(1): 1-5 | | [Pubmed] | |
|
|
 |
 |
|
|
|
|
|
|
Article Access Statistics | | Viewed | 2014 | | Printed | 61 | | Emailed | 0 | | PDF Downloaded | 116 | | Comments | [Add] | | Cited by others | 3 | |
|

|