ORIGINAL ARTICLE  
Year : 2012  |  Volume : 46  |  Issue : 1  |  Page : 46-53
Anthropometric measurements to design best-fit femoral stem for the Indian population


1 Department of Mechanical Engineering, Indian Institute of Technology, Delhi, India
2 Department of Orthopaedics, All India Institute of Medical Sciences, New Delhi, India

Click here for correspondence address and email

Date of Web Publication10-Jan-2012
 

   Abstract 

Background: The standard commercially available marketed prostheses sometimes may not be the best fit to Indian patients because of the large anatomic variation. Orthopedic surgeons always stress the need for a proper implant-patient match in hip joint replacements, in particular, for a cementless femoral stem. The complications of mismatch are aseptic loosening, improper load distribution, and discomfort. The present study was undertaken to compare the differences in dimensions between femurs of elderly Indians and those of populations from other regions in order to solve the problem of a possible geometric mismatch between a selected implant and the hip joint as far as Indian patients are concerned.
Materials and Methods: Measurements were made using computer aided design techniques on computed tomography (CT) scanned images of 98 femurs (56 left and 42 right). The software used to convert the CT images into solid models was MIMICS® (Materialize, Inc., Leuven, Belgium). The geometrical parameters, viz., the femoral head offset, femoral head center (HC), femoral head diameter, femoral head relative position, position of shaft isthmus, neck-shaft angle, bow angle, femoral neck length, canal flare index, femoral length, and canal width at various locations, were chosen to design best-fit standard femoral stems for cementless insertion. These data were compared with the published data of other countries.
Results:
A difference of 16.8% was found in the femoral head offset between Indian and Swiss populations, which can affect soft tissue tension and range of motion. At a distance of 20 mm above the lesser trochanter (LT), the anteroposterior (AP) canal width was found to differ by 45.4%, when compared with a French population which can affect the mechanical stability of femoral stem. Femoral dimensions of Indian male and female subjects have also been compared and differences evaluated. At the LT, the aspect ratio (ratio of mediolateral canal width and AP canal width) in case of males (1.198) is approximately13% higher than that of females (1.059).
Conclusions:
This study indicates a need for redesign of femoral stems. The obtained anthropometric femoral dimensions can be used to design and develop indigenous hip joint prosthesis in India. The results of this study can also be used in forensic anthropometric studies.

Keywords: Cementless femoral stem, implant design, implant-patient match, Indian population

How to cite this article:
Rawal B R, Ribeiro R, Malhotra R, Bhatnagar N. Anthropometric measurements to design best-fit femoral stem for the Indian population. Indian J Orthop 2012;46:46-53

How to cite this URL:
Rawal B R, Ribeiro R, Malhotra R, Bhatnagar N. Anthropometric measurements to design best-fit femoral stem for the Indian population. Indian J Orthop [serial online] 2012 [cited 2018 Dec 13];46:46-53. Available from: http://www.ijoonline.com/text.asp?2012/46/1/46/91634

   Introduction Top


The most common cause of hip joint failure is osteoarthritis. This hip joint failure is due to the damage of hip joint mechanism which affects range of motion and ability to bear weight on the joint. Other conditions that lead to hip replacement surgery are rheumatoid arthritis, osteonecrosis, trauma and bone tumors. [1] The most common method of treating a hip joint failure is total hip arthroplasty (THA). The prosthetic components used in THA are made of metals, polymers or ceramics, leading to improved mobility and relief of pain. [2] Over 800,000 artificial hip joint replacements are done annually worldwide. [3]

It is vital to match the dimensions of the implant closely with those of the femur, [4] as some of the complications resulting from mismatch could be aseptic loosening, improper load distribution and discomfort. Secondary biologic fixation of a hip implant depends to a large extent on the quality of its primary stability. [5],[6],[7] A mismatch in dimensions between the femur bone and prosthesis leads to micromotion of the implanted stem during the early days of post surgery, which hinders the ingrowth of trabeculae bone. It is also vital to design a prosthesis through which adequate loads can be transferred to the bone, preventing stress shielding. [8] The geometry of the proximal femur is determined by genetic and environmental factors such as age, race, sex and lifestyle. [9],[10] A population based study was carried out by Nurzenski et al. who found that lifestyle factors influence geometric indices of bone strength in the proximal femur. [11]

Siwach et al. [12] compared the parameters of the femurs of Indian cadavers with those of Western, and Hong Kong Chinese populations. They observed that the implants were oversized, and their angles and orientations were also having a mismatch which can presumably lead to complications like splintering and fractures. Reddy et al. [13] highlighted that a mismatch between femoral bone and stem definitely results in micromotion. These micromotions eventually lead to thigh pain, osteolysis and aseptic loosening. [13] Leung et al. [14],[15],[16] were prompted to modify the gamma nail (used for fixing a femoral neck) to suit the Asian population. Some researchers also investigated the differences in femoral bone parameters between male and female femurs. [17],[18]

There has long been a belief among Indian and Asia-Pacific arthroplasty surgeons that the prosthetic components currently available on the market do not fulfill the requirements of these anthropometrically smaller ethnic groups, especially in the smaller sizes. [19] If the implant is too large, the femur can fracture as it is driven down inside the bone, so the tendency is to undersize for safety. But if the implant is highly undersized, the bone may fail to bond to it. So, the correct implant size is very important. [20] A similar study was carried out by Khang et al. [21] to investigate the anatomic geometric differences between femurs from Korean subjects and those of American and Japanese subjects, and they suggested to design a new hip prosthesis system for Korean, Japanese, and other Asian patients.

Comparisons of skeletal geometric features that confer hip implant fitment between race and ethnic groups may yield insights about the mechanisms of hip implant fitment that could contribute to design a best fit hip implant among older Indians. The present study was undertaken to compare the differences in dimensions between femurs of elderly Indian as measured on CT scan images and those of populations of other region, as separate templates and design may be required for different regions.


   Materials and Methods Top


Patients and scan details

Patients were enrolled irrespective of age, sex, and region. All patients had some or other problems related to hip joint, like osteoarthritis, pain, fracture, etc. Patients who had problems in areas other than the hip joint were not selected for CT scan.

Dimensions were extracted from 56 left and 42 right femurs which belonged to 29 female and 31 male subjects (both femurs of a few patients [n=28] and a single femur of the remaining patients [n=42]). The mean age of the total population was 61.3 years (range 40-81 years). This sample population included patients from different regions of India. CT scanned images were obtained in Digital Imaging and Communications in Medicine (DICOM) format. The slice thickness of the scans was 1.25 mm. The patients were scanned for CT in a supine position with neutral rotation of the lower limb. The scanner in use was an LX Horizon high-resolution model (GE, Milwaukee, WI, USA). As a prerequisite, consent of patients was obtained for taking CT scan for study as per the guidelines of the ethics committee.

Measurements of various anthropometric parameters of an Indian population were obtained from computed tomography (CT) scanned images, using computer-aided design (CAD) techniques. It has been found that measurement from 3D images obtained through CT scans are more accurate and easier to obtain than from other methods such as 2D radiographs and direct measurement of cadaveric bones. [22],[23] These Indian femur bone values were compared with the published values of Swiss, [22] French, [5] Thai [24] and Caucasian [25] populations. A statistical analysis was conducted to determine significant differences in parametric values between the Indian population and that of other regions. These anthropometric differences, along with the required range of motion related to the Indian lifestyle (e.g. the squatting position), call for the development of a modified hip joint prosthesis for the Indian population.

Measurement of studied parameters

The CT scan data were imported into a 3D imaging software to obtain a 3D graphic model by using region growing and thresholding techniques. After geometrical simplification of the 3D model of the femur, the dimensions of studied parameters were measured in true 2D and 3D models. Most of the measurements in this study were based on an advanced method of using CT images combined with 3D reverse engineering, and CAD techniques. A graphic model of each femur was obtained by thresholding and region growing techniques. This methodology has been outlined in a reference by Mahasavairiya et al. [24]

The following parameters of the femur were chosen to be extracted for understanding the anthropometry and eventually designing a best-fit standard femoral stem for cementless insertion, as shown in [Figure 1] and [Figure 2]:

  • Femoral head offset (A)
  • Femoral head center (HC) and femoral head diameter (B)
  • Femoral head relative position (C)
  • Position of shaft isthmus (I)
  • Neck-shaft angle (J) and bow angle (K)
  • Femoral neck length (P)
  • Canal flare index (Q) and Anteversion angle (R), methodology has been outlined in the reference [25]
  • Femoral length (S), methodology has been outlined in the reference. [26]
  • Canal width (D, E, F, G, H, and L, M, N, O)
Figure 1: Femur model in polylines showing neck and femoral axes

Click here to view
Figure 2: Measurement of femoral head offset (A), femoral head diameter (B), and femoral head position (C), isthmus position (I), neck-shaft angle (J), bow angle (K), femoral neck length (P) and femoral lengths (S)

Click here to view


Canal width

The snapshot of extracted mediolateral (ML) measured dimensions and anteroposterior (AP) measured dimensions of an average elderly Indian femur are shown in [Figure 3]a and b. Sample values have been indicated there upon.

  • D: ML canal width, 20 mm above the lesser trochanter (LT)
  • E: ML canal width at the level of the LT
  • F: ML canal width, 20 mm below the LT
  • G: ML canal width at the isthmus
  • H: Periosteal (cortex + canal width) width of femur at the isthmus
  • L: AP canal width, 20 mm above the LT
  • M: AP canal width at the level of the LT
  • N: AP canal width, 20 mm below the LT
  • O: AP canal width at the isthmus
Figure 3: (a) Mediolateral measured dimensions; (b) anteroposterior measured dimensions

Click here to view
Relative positions at 20 mm above and below the level of LT were chosen as this is a standard practice for measuring hip anthropometric data and then these data were relatively compared with other studies [5],[22],[25],[26],[27],[28],[29] for each geometrical dimension which might affect the design of a femoral stem.

Statistical analysis

In the statistical analysis, the data of various measured parameters were tabulated with the mean, standard deviation, and range of observations. The data analysis was performed by using the t-test (paired) for normal distributions. A P value of less than 0.05 was considered to be significant.


   Results Top


[Table 1] compares the extracted anthropometric data of this study with the corresponding published values of other regions of the world (Swiss, French, Thai and Caucasian) and [Table 2] shows the mean values of femoral parameters for male and female subjects separately.
Table 1: Summary of the morphometry of the proximal femur reported in different studies (n = number of specimens)

Click here to view
Table 2: Femoral measurements - mean, standard deviation (SD) with range for Indian female and male specimens

Click here to view


AP and ML dimensions were measured, and AP/ML ratios were obtained and plotted. [Figure 4] illustrates the variation of the ML canal width with the AP canal width, at specific positions along the femurs, for different categories of subjects. The graph in [Figure 5] illustrates the aspect ratio for different sample categories, at different positions along the femur. The AP and ML dimensions of the femur decide the optimal AP and ML coverage of bone surface for best fit, and variations in AP and ML dimensions decide the proper size of the femoral stem.
Figure 4: Average ML canal width versus AP canal width at different positions as indicated

Click here to view
Figure 5: Comparison of mean aspect ratio at different positions

Click here to view


The graph in [Figure 6] indicates the variation of the femoral head position relative to the LT, with the femoral head offset, based on gender. This variation in femoral head position with the femoral head offset indicates the correct location of hip center (HC) and HC helps to maintain proper leverage for the soft tissues.
Figure 6: Graph showing variation of the femoral head center (HC) for male and female subjects

Click here to view



   Discussion Top


Proper sizing and placement of the prosthetic components are crucial to the success and long-term survival of a total hip arthroplasty (THA). [28] Indians and Chinese, as a part of the Asian subpopulation, have a smaller build and stature as compared with the Western population. [29] Due to the large variability of the anthropometry of different populations of the world, ethnic groups having a smaller build, such as Indians, are likely to develop technical errors in THA with most of the commercially available prostheses because of the nonavailability of smaller and proper sized implants. Studies on anthropometric measurements of the proximal femur of the average Indian population by CT scan may appear to support this argument.

[Table 1]compares the extracted anthropometric data of Indian population with the corresponding published values of other regions of the world. The methods used in such studies were radiographic for the French [5] and Caucasian [27] populations. For the Thai [24] and Swiss [22] populations, CT scans were used. Rubin et al. [22] also obtained the measurements using radiographic and direct methods and found that the mean difference obtained using radiography compared to direct measurements was 2.4 ± 1.4 mm (mean ± SD), while the difference obtained using CT scans was 0.8 ± 0.7 mm (mean ± SD). The magnitude of these errors was much lower than the measured readings. Therefore, the compared values from this study and other studies (involving radiographic measurements) can still convey a fairly accurate idea regarding the anthropometric differences. The difference in femoral head offset between Indian and Swiss populations [22] was found to be 16.8% which indicates that this can cause greater tension in soft tissues of the joint and can also increase the chances of dislocation post surgery. The ML canal width 20 mm above the LT was found to have a difference of 40% compared to the Caucasian population [27] indicating the oversize of stem for the Indians. The AP canal width 20 mm above the LT was found to have a maximum difference of 45.4% when compared with the French population [5] and this can impact the close fitting of stem in proximal part of femur, which can result in micromotion and instability. The bow angle showed a maximum difference of 10.4% compared to that of the Thai population, while the canal flare index showed a maximum percentage difference of 11% when compared with that of the Swiss population indicating a deviation in the use of cementless femoral stems for cementless type fixation. The anterior bow of the midportion of the femur is well recognized and has even been built into some current prostheses. The posterior bow of the proximal femur is just as constant as the midportion anterior bow, but it seems to have been unrecognized or is considered of no importance by most designers of femoral stems. Excessive distal femoral bowing in a lateral plane has an impact on the use of a stemmed femoral prosthesis. The cementless stem with a long bowed, distally flexible stem may be associated with greater implant stability than a more rigid, long bowed stem that requires over-reaming to avoid fracture. [30] Canal flare index for Indian population is 4.23 (mean), which is above 3 and below 5, and indicates the standard cementless femoral stem as the best choice as per Fessy et al.[31] algorithm for the choice of femoral implant.

The above differences in dimensions are mostly in the superior and proximal regions of the femora. These can, therefore, significantly affect the performance of a standard size femoral stem for cementless fixation, as pointed out by Hua and Walker. [32] Micromotion hinders bone ingrowth, thereby effecting secondary stability over a period of time. Also, the load distribution would not be well distributed, leading to early fatigue failure and possible breakage of the stem which has been echoed by Ducheyne et al.[33] in 2004.

[Table 2]shows the mean values of femoral dimensions for the male and female subjects separately. Mean values for male subjects were found to be higher for all the data except the anteversion angle (R). This could be attributed to the hip size and load distribution differences between the males and females. The neck of the femur forms a shallower angle with the long axis of the femur bone in a female. Forensic anthropologists have also found differences in the measurements of the transverse width and length of the head of the femur between the sexes and have postulated this application in identifying sex of dismembered murder victims; however, there are significant differences between races also. [9]

At the isthmus, values for both males and females are the lowest, with the AP value being larger than the ML value [Figure 4]. The values for all the categories are almost equal at this position. From the isthmus, in the proximal direction of the femur, the ML and AP canal width values increase almost linearly. It is also observed that corresponding values for different categories have mean differences of 9.5% in the proximal position. The design of a stem must provide an initial or primary stability and it should be taken into account during the design evolution stage itself. [34]

At the isthmus and 20 mm below the LT, there is not much variation in the aspect ratio among different categories [Figure 5]. At the LT, the aspect ratio (ML/AP) in case of males (1.198) is approximately 13% higher than that of females (1.059). At 20 mm above the LT, the aspect ratio for females is higher (1.458) as compared to males (1.355) by approximately 8%. This might make the matching of a standard commercially available cementless femoral stem in proximal area of femur both in male and female patients more difficult.

There are three key extramedullary dimensions, viz., leg length (or femoral head position), femoral offset, and anteversion angle, which help in maintaining proper hip kinematics and improving overall patient satisfaction post surgery of THA. [35],[36] Optimizing these parameters before actual implantation can lead to correct leg length, range of motion, and joint stability. From [ 6] it is observed that the female femoral HC lies inferiorly (related to the LT), as compared to the male femoral HC. The range of femoral head offset for females of the Indian population is found to be smaller by 37% as compared to males of the same population. The linear fit to the male and female data is approximately 10 mm apart. These data reveal that there should be relative degree of difficulty in fixing the same femoral stem to a male and female patient during THA to restore the natural mechanics of the joint by considering both extramedullary and intramedullary parameters of the femur.

The findings of this study show higher values for some major femoral dimensions when compared with international published literature. [5],[22],[23],[24] However, these values are found to be very close when compared with the other Indian studies. [12],[37],[38] The above findings indicate marked differences in the observed dimensions of the femoral canal and femoral head position between males and females. This information is vital to a mechanical/biomedical designer for creating a better design of femoral stems for cementless hip joint prostheses. These differences also influence the fit of the stems to the patient of a particular geographic region. Therefore, for best-fit cementless femoral stem design, the study suggests ML, AP and some extrafemoral dimensions as the criteria to design gender specific proper stems suitable for most of the Indian population. To illustrate this better, three groups were created based on the femoral head offset, neck-shaft angle, canal ML and AP dimensions; the smallest 25%, the middle 50%, and the largest 25% for each gender, respectively. Based on the data, we have proposed a design of three groups of femoral stem for male and female, respectively [Table 3].
Table 3: Proposed dimension of the cementless femoral stem (in mm) for the Indian population

Click here to view



   Conclusions Top


The results of this study indicate that marked differences do exist in the dimensions between the femur of the Indian population and that of the populations of other regions of the world. Canal flare index for the Indian population indicates the standard cementless femoral stem as the best choice.

There are significant differences in dimensions between male and female femora within the Indian population, indicating that a range of femoral stem designs are required to reduce the inventory and narrow down the best fit options for a surgeon. These ranges of evolved femoral stem geometries would eventually improve the clinical outcomes, reduce the cost to the patient, and eliminate the possibility of a revision surgery.

 
   References Top

1.Amstutz HC. Complications of total hip replacement. Clin Orthop 1970;72:123-37.  Back to cited text no. 1
[PUBMED]    
2.Gomez PF, Morcuende JA. Early attempts at hip arthroplasty- -1700s to 1950s. Iowa Orthop J 2005;25:25-9.  Back to cited text no. 2
[PUBMED]  [FULLTEXT]  
3.Granger C, Schutte HD, Bigger SB, Kennedy JM, Latour RA. Failure analysis of composite femoral component for hip arthoplasty. J Rehab Res Develop 2003;40:131-46.  Back to cited text no. 3
    
4.Engh CA. Hip arthroplasty with a Moore prosthesis with porous coating: A five year study. Clin Orthop 1983;176:52-66.  Back to cited text no. 4
[PUBMED]    
5.Husmann O, Rubin PJ, Leyvraz PF, de Roguin B, Argenson JN. Three-dimensional morphology of the proximal femur. J Arthroplasty 1987;12:444-50.  Back to cited text no. 5
    
6.Engh CA, Bobyn JD, Glassman AH. Porous-coated hip replacement: The factors governing bone ingrowth, stress shielding, and clinical results. J Bone Joint Surg 1987;69:45-55.  Back to cited text no. 6
    
7.Morscher E. Cementless total hip arthroplasty. Clin Orthop 1983;181:76-91.  Back to cited text no. 7
[PUBMED]    
8.Jasty M, Henshaw PM, O'Connor DO. Strain alterations in the proximal femur with an uncemented femoral prosthesis emphasizing the effect of component fix. Atlanta, GA: Proc Orthopaedic Research Society meeting; 1988. p. 335.  Back to cited text no. 8
    
9.Najjar EI, McWilliams ER. Forensic anthropology: The structure, Morphology and variations of human bone and dentition, Springfield, IL: Charles C Thomas; 1978.  Back to cited text no. 9
    
10.Ericksen MF. Ageing changes in the medullary cavity of the proximal femur in American black and whites. Am J Phys Anthropol 1979;51:563-9.  Back to cited text no. 10
[PUBMED]    
11.Nurzenski MK, Briffa NK, Price RI, Khoo BC, Devine A, Beck TJ, et al. Geometric indices of bone strength are associated with physical activity and dietary calcium intake in healthy older women. J Bone Miner Res 2007;22:416-24.  Back to cited text no. 11
[PUBMED]  [FULLTEXT]  
12.Siwach RC, Dahiya S. Anthropometric study of proximal femur geometry and its clinical application. Indian J Orthop 2003;37:247-51.  Back to cited text no. 12
  Medknow Journal  
13.Reddy VS, Moorthy GV, Reddy SG. Do we need a special design of femoral component of total hip prosthesis in our patients? Indian J Orthop 1999;33:282-4.  Back to cited text no. 13
  Medknow Journal  
14.Leung K, Procter P, Robioneck B, Behrens K. Geometric mismatch of the gamma nail to the Chinese femur. Clin Orthop 1996;323:42-8.  Back to cited text no. 14
    
15.Leung KS. Early experience with gamma nails in the treatment of peritrochanteric fractures. Trans Hong Kong Orthop Assoc 1989;33. Available from: http://www.ijoonline.com/text.asp? [Last accessed on 2011 Apr 20].  Back to cited text no. 15
    
16.Leung KS, So WS, Shen WY, Hui PW. Gamma nails and dynamic hip screws for peritrochanteric fractures. J Bone Joint Surg 1992;74:345-51.  Back to cited text no. 16
    
17.Yoshioka Y, Siu D, Cooke TD. The anatomy and functional axes of the femur. J Bone Joint Surg Am 1987;69:873-80.  Back to cited text no. 17
[PUBMED]  [FULLTEXT]  
18.Maruyama M, Feinberg JR, Capello WN, D'Antonio JA. Morphological features of the acetabulum and femur, anteversion angle and implant positioning. Clin Ortho Rel Res 2001;393:52-65.  Back to cited text no. 18
    
19.Vaidya SV, Ranawat CS, Aroojis A, Laud NS. Anthropometric measurements to design total knee prostheses for the Indian Population. J Arthroplast 2000;15:79-85.  Back to cited text no. 19
    
20.Available from: http://hipsandknees.com/hip/hipimplants.htm [Last cited on 2011 May 11].  Back to cited text no. 20
    
21.Khang G, Choi K, Kim C, Yang J, Bae T. A study of Korean femoral geometry. Clin Orthop Relat Res 2003;406:116-22.  Back to cited text no. 21
    
22.Rubin PJ, Leyvraz PF, Aubagniac JM. The morphology of the proximal femur: A three-dimensional radiographic analysis. J Bone Joint Surg 1992;74:28-32.  Back to cited text no. 22
    
23.Rubin PJ, Leyvraz PF, Heegaard JH. Radiological variations in the anatomical parameters of the proximal femur in relation to rotation. Fr J Orthop Surg 1989;3:121.  Back to cited text no. 23
    
24.Mahaisavariya B, Sitthiseripratip K, Tongdee T, Bohez EL, Sloten JV, Oris P. Morphological study of the proximal femur: A new method of geometrical assessment using 3-dimensional reverse engineering. Med Engg Phys 2002;24:617-22.  Back to cited text no. 24
    
25.Weiner DS, Cook AJ, Hoyt WA, Oravec CE. Computed tomography in the measurement of femoral anteversion. Orthopedics 1987;1:299-306.  Back to cited text no. 25
    
26.Luo W, Stanhope SJ, Sheehan FT. Using two palpable measurements improves the subject-specific femoral modeling. J Biomech 2009;42:2000-5.  Back to cited text no. 26
[PUBMED]  [FULLTEXT]  
27.Noble PC, Jerry W, Alexander JW, Lindhal LJ, Yew DT, Granberry WM, et al. The anatomical basis of femoral component design. Clin Orthop 1988;235:148-65.  Back to cited text no. 27
    
28.Byung WN, Kwang SS, Ki CB, Chul HC, Chul HK, Shin YK. The effect of stem alignment on results of total hip arthroplasty with a cementless tapered-wedge femoral component. J Arthroplast 2008;23:418-23.  Back to cited text no. 28
    
29.Hoaglund FT, Low WD. Anatomy of the femoral neck and head with comparative data Caucasians and Hong Kong Chinese. Clin Orthop 1980;152:10.  Back to cited text no. 29
[PUBMED]    
30.Ries MD, Suzuki Y, Renowitzky G, Lotz JC, Barrack RL, Bourne RB, et al. Effect of cementless bowed stem distal surface contour and coronal slot on femoral bone strains and torsional stability. J Arthroplasty 2003;18:494-8.  Back to cited text no. 30
[PUBMED]  [FULLTEXT]  
31.Fessy MH, Seutin B, Bdjui J. Anatomical basis for the choice of the femoral implant in the total hip arthroplasty. Surg Radiol Anat 1997;19:283-6.  Back to cited text no. 31
    
32.Hua J, Walker PS. Closeness of fit of uncemented stems improves the strain distribution in the femur. J Orthop Res 1995;13:339-46.  Back to cited text no. 32
[PUBMED]    
33.Ducheyne P, Aernoudt E, Meester PD, Martens M, Mulier JC, Leeuwen DV. Factors governing the mechanical behavior of the implant-porous coating-trabecular bone interface. J Biomech 2004;19:658-60.  Back to cited text no. 33
    
34.Fuku H. Cementless hip prosthesis design: A basic study and analysis of the proximal femur in normal Japanese people. Nippon Seikeigeka Gakkai Zasshi 1994;68:763-73.  Back to cited text no. 34
[PUBMED]    
35.Sugano N, Noble PC, Kamaric E. Predicting the position of the femoral head center. J Arthroplasty 1999;14:102-7.   Back to cited text no. 35
[PUBMED]  [FULLTEXT]  
36.Ranawat CS, Rodriguez JA. Functional leg-lenth inequality following total hip arthroplasty. J Arthroplasty 1997;12:359-64.  Back to cited text no. 36
[PUBMED]    
37.Maheshwari AV, Zlowodzki MP, Siram G, Jain AK. Femoral neck anteversion, acetabular anteversion and combined anteversion in the normal Indian adult population: A computed tomographic study. Indian J Orthop 2010;44:277-82.  Back to cited text no. 37
[PUBMED]  Medknow Journal  
38.Jain AK, Maheshwari AV, Nath S, Singh MP, Nagar M. Anteversion of the femoral neck in Indian dry femora. J Orthop Sci 2003;8:334-40.  Back to cited text no. 38
[PUBMED]  [FULLTEXT]  

Top
Correspondence Address:
B R Rawal
Department of Mechanical Engineering, Indian Institute of Technology Delhi, New Delhi
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0019-5413.91634

Rights and Permissions


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6]
 
 
    Tables

  [Table 1], [Table 2], [Table 3]

This article has been cited by
1 Femoral Head Fovea Capitis Variant Configurations and Age Related Changes- A Radiological Study
Nihat Acar,Ahmet Karaarslan,Ahmet Karakasli,Mehmet Erduran
Iranian Journal of Radiology. 2017; InPress(InPress)
[Pubmed] | [DOI]
2 Additive Manufacturing Fixture Box for Bone Measurement
V. Sindhu,S. Soundarapandian
Procedia Engineering. 2017; 184: 1
[Pubmed] | [DOI]
3 A parametric approach to construct femur models and their fixation plates
Xiaozhong Chen,Kunjin He,Zhengming Chen,Lin Wang
Biotechnology & Biotechnological Equipment. 2016; 30(3): 529
[Pubmed] | [DOI]
4 Exploring inter-subject anatomic variability using a population of patient-specific femurs and a statistical shape and intensity model
Mamadou T. Bah,Junfen Shi,Martin Browne,Yanneck Suchier,Fabien Lefebvre,Philippe Young,Leonard King,Doug G. Dunlop,Markus O. Heller
Medical Engineering & Physics. 2015; 37(10): 995
[Pubmed] | [DOI]
5 Three-Dimensional Analysis of the Characteristics of the Femoral Canal Isthmus: An Anatomical Study
Xiu-yun Su,Jing-xin Zhao,Zhe Zhao,Li-cheng Zhang,Chen Li,Jian-tao Li,Jian-feng Zhou,Li-hai Zhang,Pei-fu Tang
BioMed Research International. 2015; 2015: 1
[Pubmed] | [DOI]
6 Femoral Version, Neck-Shaft Angle, and Acetabular Anteversion in Chinese Han Population
Nan Jiang,Lin Peng,Mohammed Al-Qwbani,Guo-Ping Xie,Qin-Meng Yang,Yu Chai,Qing Zhang,Bin Yu
Medicine. 2015; 94(21): e891
[Pubmed] | [DOI]
7 Proximal Femoral Morphology and the Relevance to Design of Anatomically Precontoured Plates: A Study of the Chinese Population
Kun-Jhih Lin,Hung-Wen Wei,Kang-Ping Lin,Cheng-Lun Tsai,Pei-Yuan Lee
The Scientific World Journal. 2014; 2014: 1
[Pubmed] | [DOI]
8 Morphological Study of the Newly Designed Cementless Femoral Stem
Mohd Yusof Baharuddin,Sh-Hussain Salleh,Ahmad Hafiz Zulkifly,Muhammad Hisyam Lee,Alias Mohd Noor
BioMed Research International. 2014; 2014: 1
[Pubmed] | [DOI]
9 The effects of age and gender on the diameter of the femoral canal in patients who undergo total hip replacement
D. J. Milligan,S. OæBrien,D. Bennett,J. C. Hill,D. E. Beverland
The Bone & Joint Journal. 2013; 95-B(3): 339
[Pubmed] | [DOI]
10 Are the Left and Right Proximal Femurs Symmetric?
Ernest Y. Young,Jeremy Gebhart,Daniel Cooperman,Nicholas U. Ahn
Clinical Orthopaedics and Related Research®. 2013; 471(5): 1593
[Pubmed] | [DOI]
11 Radiographic Fit and Fill Analysis of a New Second-Generation Proximally Coated Cementless Stem Compared To Its Predicate Design
Kimona Issa,Robert Pivec,Thies Wuestemann,Tiffany Tatevossian,Jim Nevelos,Michael A. Mont
The Journal of Arthroplasty. 2013;
[Pubmed] | [DOI]
12 Multi-pelvis characterisation of articular cartilage geometry
Gillard, F.C., Dickinson, A.S., Schneider, U., Taylor, A.C., Browne, M.
Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 2013; 227(12): 1255-1264
[Pubmed]
13 Three dimensional morphometry of the femur to design the total hip arthroplasty for Malay population
Baharuddin, M.Y. and Zulkifly, A.H. and Lee, M.H. and Kadir, M.R.A. and Saat, A. and Aziz, A.A.
Advanced Science Letters. 2013; 19(10): 2982-2987
[Pubmed]
14 Are the left and right proximal femurs symmetric? Basic research
Young, E.Y. and Gebhart, J. and Cooperman, D. and Ahn, N.U.
Clinical Orthopaedics and Related Research. 2013; 471(5): 1593-1601
[Pubmed]
15 The effects of age and gender on the diameter of the femoral canal in patients who undergo total hip replacement
Milligan, D.J. and OæBrien, S. and Bennett, D. and Hill, J.C. and Beverland, D.E.
Journal of Bone and Joint Surgery - Series B. 2013; 95 B(3): 339-342
[Pubmed]
16 Significance of canal flare index measurement in total hip arthroplasty
Li, Y.-Z., Li, J.-L., Lin, J.-K.,Pan, Y.-C., Guo, L.-R.
Chinese Journal of Tissue Engineering Research. 2012; 16(52): 9702-9706
[Pubmed]



 

Top
 
 
 
  Search
 
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Email Alert *
    Add to My List *
* Registration required (free)  
 


 
    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusions
    References