CASE REPORT  
Year : 2012  |  Volume : 46  |  Issue : 1  |  Page : 109-112
Broken guidewire protruding into the hip joint: A bone endoscopic-assisted retrieval method


Department of Orthopedic Surgery, Maulana Azad Medical College and Associated Lok Nayak Hospital, New Delhi, India

Click here for correspondence address and email

Date of Web Publication10-Jan-2012
 

   Abstract 

Broken implants, especially broken wires at difficult sites, may pose a challenge for the treating orthopedic surgeon. We describe a method for extraction of a broken guidewire that was, protruding into the hip joint following the insertion of a proximal femoral nail. A 35-year-old man with displaced femoral neck fracture with ipsilateral fracture shaft of femur was operated and fixed with long proximal femoral nail. The guidewire of proximal screw broke during the process of drilling. The tip of the 2-cm-long broken guidewire was touching the articular surface. The guidewire was misdirected posteroinferiorly from its path for the insertion of the proximal screw (6.8 mm), this screw was removed and bone endoscopy was performed with a 30° arthroscope. The broken end of the guidewire was located under direct vision. The grasper was introduced with its jaws at the 8 O'clock position and its position was confirmed under a C-arm image intensifier in both anteroposterior and lateral views. The broken end of the guidewire was grasped and it was retrieved. The screw was replaced in its original track to complete the procedure. The fractures united and patient was asymptomatic when last followed-up at 12 months.

Keywords: Bone endoscopy, broken guidewire, proximal femoral nail, femoral neck fracture

How to cite this article:
Arora S, Maini L, Aggarwal V, Dhal A. Broken guidewire protruding into the hip joint: A bone endoscopic-assisted retrieval method. Indian J Orthop 2012;46:109-12

How to cite this URL:
Arora S, Maini L, Aggarwal V, Dhal A. Broken guidewire protruding into the hip joint: A bone endoscopic-assisted retrieval method. Indian J Orthop [serial online] 2012 [cited 2019 Dec 12];46:109-12. Available from: http://www.ijoonline.com/text.asp?2012/46/1/109/91646

   Introduction Top


A broken wire within a bone usually does not warrant removal. Rarely a wire may break inside the bone and protrude into the neighboring joint, necessitating its removal in order to avoid damage to articular cartilage and consequent early degenerative arthritis. Several techniques have been described for the removal of a broken intramedullary nail. [1],[2],[3],[4],[5] Bone endoscopy (actually medulloscopy) has been used for removal of a broken intramedullary nail, assisted closed reduction of long bone fractures, intramedullary loss of reamer, cement removal in revision hip arthroplasty, curettage in simple bone cyst, and direct visual confirmation of cannulated screw placement in slipped capital femoral epiphysis. [6],[7],[8],[9],[10],[11] Here, we describe a method for removal of a broken guidewire in the hip joint. To the best of our knowledge, a bone endoscopic-assisted method for removing a broken guidewire has not been described before in the English-language literature.


   Case Report Top


A 35-year-old man presented to the orthopedic emergency with the history of road traffic accident a few hours back. He was diagnosed as closed, short oblique fracture of the mid shaft of the right femur, with a small butterfly fragment (involving less than 25% of its circumference), and ipsilateral displaced femoral neck fracture. Skeletal traction through an upper tibial pin was applied pending operative fixation of both the fractures.

A closed reduction and fixation of the shaft as well as the neck of femur was performed using a long proximal femoral nail (Yogeshwar Pvt. Ltd, Mumbai, India). During surgery, the guidewire of the proximal-most screw of the neck of femur broke during the process of drilling and became misdirected posteroinferiorly. Anteroposterior, lateral, and oblique views taken under a C-arm image intensifier showed that the tip of the 2-cm-long broken guidewire was just touching the articular surface of the femoral head, without any protrusion into the hip joint. The surgical procedure was completed in the usual manner. The postoperative period was uneventful, although patient did complain of hip pain. Clinical examination failed to elicit crepitus in the hip joint. As the postoperative radiographs were equivocal [Figure 1], fine-cut multidetector computed tomography of the involved hip was performed, which suggested a 3-mm wire-tip protrusion into the hip joint [Figure 2].
Figure 1: Postoperative radiographs (anteroposterior and lateral view) showing the long proximal femoral nail and the broken guidewire near the lower margin of the proximal-most screw

Click here to view
Figure 2: 3 mm cuts of computed tomography, coronal images, showing wire-tip protrusion into the hip joint

Click here to view


It was necessary to preempt the possibility of articular cartilage abrasion by the protruding wire, and so the various options considered for the removal of offending wire were (1) arthroscopic extraction, (2) arthroscopic push down into the femoral head, or (3) bone endoscopy-assisted extraction. We attempted the bone endoscopy-assisted method for extraction of the broken wire. Since the guidewire was misdirected posteroinferiorly from its path for the insertion of the proximal screw (6.8 mm), this screw was removed and bone endoscopy was performed with a 30° arthroscope [Figure 3]a. The broken end of the guidewire was located under direct visualization [Figure 3]b. The grasper was introduced with its jaws at the 8 O'clock position and its position was confirmed under a C-arm image intensifier in both anteroposterior and lateral views [Figure 4]a. The broken end of the guidewire was grasped and it was retrieved [Figure 4]b The screw was replaced in its original track to complete the procedure. The screw purchase was found to be satisfactory.
Figure 3: a: Fluoroscopic view showing insertion of scope into the screw tunnel
Figure 3: b: Bone endoscopic view of the tunnel showing the broken wire posteroinferiorly


Click here to view
Figure 4: a: Fluoroscopic view showing the grasper being advanced in the direction of the visualized misplaced broken guidewire and the jaws were opened after directing them towards the broken guidewire
Figure 4: b: The retrieved guidewire (approximately 2 cm), along with the grasper used to retrieve it
Figure 4: c: Follow-up radiograph (anteroposterior view) showing union of both the fractures


Click here to view


The patient was started on quadriceps exercises and active knee bending exercises 3 days after surgery. As the patient's comfort increased, he was mobilized without weight bearing on the operated side and graduated to partial weight bearing at 12 weeks. Both the fractures united and patient was asymptomatic when last followed-up at 12 months [Figure 4]c.

Written informed consent was obtained from the patient. He was also informed that data concerning the case would be submitted for publication.


   Discussion Top


A broken Kirschner wire/guidewire/drill bit usually does not require removal except in exceptional circumstances, like breakage with protrusion into the joint, intra-articular migration, or compression over neurovascular structures. Removal from difficult situations such as when there is intraosseous breakage of the wire with protrusion into the joint can pose a dilemma, considering the magnitude of iatrogenic damage caused by an arthrotomy, especially in the hip joint.

Oberst et al. pioneered the technique of intramedullary bone endoscopy (IBE) or medulloscopy. [6],[7],[8],[12],[13] They first performed IBE in the cadaveric femur and tibia and then extended its use to clinical situations for removal of broken nail and assisted closed reduction of long bone fractures. They used a special endoscope with an oval working canal. Govaers et al. successfully used the method of medulloscopy for cement removal in revision arthroplasty of the hip. [10]

We considered various options for the extraction of the offending wire, including hip arthroscopic extraction, hip arthroscopic push down into the femoral head, and bone endoscopy-assisted extraction. An arthroscopic extraction of the wire would have required sufficient distraction of the joint and would have been difficult to catch hold of the small protruding wire-tip using graspers. We also considered the hip arthroscopic push down of the wire into the femoral head to be a suitable option, but engagement of threaded end of the guidewire in the subchondral bone would have posed an obstacle. Considering the magnitude of damage caused by surgical dislocation of the hip, we thought it would be imprudent to use this method for the guidewire removal. Bone endoscopy through the screw track after temporary removal of the screw was considered the most suitable option. It may be desirable to use the grasper under direct scope visualization. We were, however, restricted by the screw tract diameter, which could accommodate only one instrument at a time. Scope visualization helped in direct visual localization of the broken guidewire in the screw tract. This gave us a fairly accurate impression about its location, which was in the 8 O'clock position in the relationship to the cross-section of screw tract. This helped in the ultimate retrieval of the broken wire using the grasper, with help from the image intensifier. We strongly feel that preliminary scope visualization reduced the overall radiation exposure; had the image intensifier being the only tool for localization, there would have been greater radiation to the patient due to the repeated attempts that are usually required for retrieval of a broken guidewire. Furthermore, the procedure done without visual localization is blind procedure, with the risk of articular cartilage perforation and joint penetration by the instruments.

The advantages of this method were its minimally invasive nature and the direct visualization of the guidewire, which resulted in overall reduction in radiation exposure to surgeons as well as patient. However, the need for sophisticated instrumentation may limit its use in centers with limited resources. Removal of the screw and reinsertion into the same track after the procedure carries the inherent risk of loss of purchase and the need for an additional screw insertion. An arthroscope can be used in place of the special endoscope in selected circumstances. As in the present case, the canal (that we utilized for passing the scope) was prepared for screw insertion during the original surgery. The need for an image intensifier may be completely obviated with the use of a special endoscope with a working canal. [7]

 
   References Top

1.Whalley H, Thomas G, Hull P, Porter K. Surgeon versus metalwork- tips to remove a retained intramedullary nail fragment. Injury 2009;40:783-9.  Back to cited text no. 1
    
2.Maini L, Upadhyay A, Aggarwal A, Dhaon BK. A new method of removing a fractured interlocked nail. Injury 2002;33:261-2.  Back to cited text no. 2
    
3.Maini L, Singh J, Agarwal P, Gautam VK, Dhal AK. Removal of broken Kuntscher's nail: An innovative close technique. J Trauma 2005;59:1518-20.  Back to cited text no. 3
    
4.Maini L, Jain N, Singh J, Singh H, Bahl A, Gautam VK. Removal of a multisegmental broken nail by close technique using a TEN nail. J Trauma 2009;66:e78-80.  Back to cited text no. 4
    
5.Tadros AM, Blachut P. Segmentally fractured femoral Küntscher nail extraction using a variety of techniques. Am J Orthop (Belle Mead NJ) 2009;38:e59-60.  Back to cited text no. 5
    
6.Oberst M, Schlegel K, Mory C, Suedkamp N. Endoscopically controlled removal of a broken intramedullary nail: A new technique. Inj Extra 2005;36:582-5.  Back to cited text no. 6
    
7.Oberst M, Niemeyer P, Swedkamp N. Endoscopic controlled intramedullary fracture reduction. A case report about closed fracture reduction under visual control. Arch Orthop Trauma Surg 2007;127:647-9.  Back to cited text no. 7
    
8.Oberst M, Holz U. Intramedullary bone endoscopy. First clinical application of a novel endoscopic technique. Unfallchirurg 2005;108:75-8.   Back to cited text no. 8
    
9.Bassett GS. Bone endoscopy: Direct visual confirmation of cannulated screw placement in slipped capital femoral epiphysis. J Pediatr Orthop 1993;13:159-63.  Back to cited text no. 9
    
10.Govaers K, Meermans G, Stuyck J, Deprez P, Bortier H, Roels J. Endoscopy for cement removal in revision arthroplasty of the hip. J Bone Joint Surg Am 2006:88 (Suppl 4):101-9.  Back to cited text no. 10
    
11.Bonnel F, Canovas F, Faure P. Treatment of a simple bone cyst of the calcaneus by endoscopic curettage with cancellous bone injection. Acta Orthop Belg 1999;65:528-31.  Back to cited text no. 11
    
12.Oberst M, Bosse A, Holz U. Intramedullary bone endoscopy (IBE). First results of experimental intramedullary endoscopy of long bone. Unfallchirurg 2002;105:853-5.   Back to cited text no. 12
    
13.Oberst M, Bosse A, Holz U. Intramedullary fracture reduction of long bone under visual control. Experimental results of an endoscopic technique. Arthroscopy 2006;22:686.e1-5.  Back to cited text no. 13
    

Top
Correspondence Address:
Sumit Arora
C/o Mr. Sham Khanna, 2/2, Vijay Nagar, Delhi - 110 009
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0019-5413.91646

Rights and Permissions


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4]



 

Top
 
 
 
  Search
 
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Email Alert *
    Add to My List *
* Registration required (free)  
 


 
    Abstract
   Introduction
   Case Report
   Discussion
    References
    Article Figures
 

 Article Access Statistics
    Viewed2293    
    Printed60    
    Emailed1    
    PDF Downloaded183    
    Comments [Add]    

Recommend this journal